12. Using a high voltage probe ( RS type 610 281 ) and meter ( RS type 610
950 ) measure the voltage on each of the output sockets. The reading should
be at minimum power 4kV and at maximum power 8 kV
13. Having checked the power supply, reconnect one Model 976EX Bar.
14.
Using a high voltage probe (RS 610 281) and meter (RS 610 590)
measure the voltage on the pins of the bar. This voltage should be
between 3 and 6kV.
15.
If there is more than one bar to test, disconnect the first item and repeat the
above steps.
13