MedViron 39

4.6  

Assemble bottom washer to the new gas spring by screwing the bottom washer onto the gas 

spring. Re-attach the nut onto the bottom washer/gas spring. (Figure 7). 

4.7      Insert the gas spring into the column and attach the bottom washer to the column with the two 
screws removed in step 5.7 making sure all bolt holes line up with screw boses in the column. 

4.8     Attach the Base to the column using the screws, washers & anti-mushroom plate removed in step 
4.4 (Figure 6). 

4.9 

Attach the top washer removed in step 4.3 to top of gas spring. Thread on the top nut.  Adjust 

the nut setting so that the top of the lever is parellel to the top washer (Figure 5). 

4.10

 

Attach the upper column to the top washer.  Use the provided block to boost the upper column 

(Figure 8).  Activate the gas spring lever and push the top plate down, aligning the plate and column 
holes.  Attach with the 2 screws removed in step 4.3 (Figure 5). 

Figure 8 

 

4.11

 

Attach the Top plate to the column with the 4 screws removed in step 4.2 (Figure 4).  

 

4.12

 

Attach the top as shown in Section 3.  
 

Section 5:  Service Instructions Vanity Tray Replacement 

This manual is intended for trained professionals for adjustment of the Medviron Overbed Table.  
Any action not outlined in this manual may result in voiding of the warranty. 

Tools that may be required 

Straight or Slotted screwdriver 

#2 Phillips screwdriver 

 

5.0        Removing or replacing a vanity tray 

5.1        Removing the vanity tray 

Summary of Contents for 39" Overbed Table

Page 1: ...Service Manual 32 Medviron Overbed Table...

Page 2: ...Table of Contents Section 1 Caster Replacement Section 2 Plug Replacement Section 3 Top Replacement Section 4 Gas Spring Replacement Section 5 Vanity Service...

Page 3: ...traight slot screw driver 1 0 Turn the over bed table upside down To replace casters that are worn or malfunctioning first raise the unit to its fullest extension Turn the over bed table upside down a...

Page 4: ...e upside down or on its side To replace damaged or worn end plugs first raise the unit to its fullest extension Turn the over bed table upside down or on its side and place on a protected surface such...

Page 5: ...esult in voiding of the warranty Tools required 5 32 Allen wrench 2 Square driver bit Provided 3 0 Remove the top Raise the unit to its fullest extension Remove 8 screws from the underside of the plat...

Page 6: ...t This gas spring keeps the table from being lowered until the handle is activated but allows free rise Gas springs can lose force over time which will cause the table to descend faster when activated...

Page 7: ...gas spring to the top washer Remove the top nut the top washer set aside Figure 5 Figure 6 4 4 Remove the base Place unit on a soft surface such as a blanket or carpet Remove four bolts washers Anti...

Page 8: ...arellel to the top washer Figure 5 4 10 Attach the upper column to the top washer Use the provided block to boost the upper column Figure 8 Activate the gas spring lever and push the top plate down al...

Page 9: ...s described in 5 1 Insert the new vanity If the vanity works well and does not slide too hard or easy no adjustment is necessary If the vanity slides too hard or too easily adjustment may be required...

Reviews: