![MedViron 39](http://html1.mh-extra.com/html/medviron/39-overbed-table/39-overbed-table_service-manual_1763241008.webp)
4.6
Assemble bottom washer to the new gas spring by screwing the bottom washer onto the gas
spring. Re-attach the nut onto the bottom washer/gas spring. (Figure 7).
4.7 Insert the gas spring into the column and attach the bottom washer to the column with the two
screws removed in step 5.7 making sure all bolt holes line up with screw boses in the column.
4.8 Attach the Base to the column using the screws, washers & anti-mushroom plate removed in step
4.4 (Figure 6).
4.9
Attach the top washer removed in step 4.3 to top of gas spring. Thread on the top nut. Adjust
the nut setting so that the top of the lever is parellel to the top washer (Figure 5).
4.10
Attach the upper column to the top washer. Use the provided block to boost the upper column
(Figure 8). Activate the gas spring lever and push the top plate down, aligning the plate and column
holes. Attach with the 2 screws removed in step 4.3 (Figure 5).
Figure 8
4.11
Attach the Top plate to the column with the 4 screws removed in step 4.2 (Figure 4).
4.12
Attach the top as shown in Section 3.
Section 5: Service Instructions Vanity Tray Replacement
This manual is intended for trained professionals for adjustment of the Medviron Overbed Table.
Any action not outlined in this manual may result in voiding of the warranty.
Tools that may be required
Straight or Slotted screwdriver
#2 Phillips screwdriver
5.0 Removing or replacing a vanity tray
5.1 Removing the vanity tray