background image

Medtrum 

Simplifying Diabetes

w

w

w

.m

edt
ru

m

.com

CGM System Overview

1

Charge PDM & Transmitter

2

PDM Settings

3

CGM Alerts Settings

5

Apply Sensor

6

Start Sensor

8

Read CGM

9

Disconnect Sensor

11

Your Personal Settings

12

Contents

Summary of Contents for TouchCare

Page 1: ...www medtrum com TouchCare CGM System PDM Quick Start Guide mmol L ...

Page 2: ...lifying Diabetes www medtrum com CGM System Overview 1 Charge PDM Transmitter 2 PDM Settings 3 CGM Alerts Settings 5 Apply Sensor 6 Start Sensor 8 Read CGM 9 Disconnect Sensor 11 Your Personal Settings 12 Contents ...

Page 3: ...in to measure your glucose level in interstitial fluid The Transmitter MD1158 records Sensor data and sends data to the PDM via Bluetooth Low Energy The Sensor with Transmitter snapped in is waterproof at 8 feet 2 5 meters for 60 minutes IP28 Store the Sensor at temperatures between 2 C 36 F and 30 C 86 F Store the Transmitter at temperatures between 10 C 14 F and 55 C 131 F Glucose Sensor Transmi...

Page 4: ...sor use Do NOT store the Transmitter connected with a Sensor or a USB charging cable That may kill the transmitter battery The indicator light will flash green when the Transmitter is charging and turn off when the Transmitter is fully charged Fully charged Charging Not charging Battery low Battery empty Charge PDM The PDM requires an AC adapter with an output of DC 5 0V Charge PDM Transmitter 3 ...

Page 5: ...reen light will flash Turn off Long press the power button for about 2 seconds then slide to power off Or long press the power button for about 8 seconds 4 Set time and date 2 Wake the PDM When the screen goes black press the power button and then slide to unlock 3 Select language and country PDM Settings 4 ...

Page 6: ...tter SN Main Menu Sensor Transmitter SN Tap to enter SN manually 6 Set display settings Main Menu Settings General Display 5 Set audio options Main Menu Settings General Audio Options If you set the audio option to Audio off Vibrate off your PDM only vibrates when a serious alert occurs It is recommended for new users to set backlight time as 2 minutes PDM Settings 5 ...

Page 7: ...ff the Alert switch Tap to save the settings High Low Limits Up to 8 segments for each day Tap to save the settings Tap Add Time Segment to add a new segment Slide to left on each segment tap Delete The start time of next segment will be adjusted Slide on segment back to right to cancel deleting Predictive Alerts You will be reminded of a predicted high or low glucose value some time the predictiv...

Page 8: ...ected insertion area with rubbing alcohol and wait for the area to dry up The Sensor adhesive stays more firmly on dry skin Place the Sensor Use your thumb and finger to bend off the safe lock by pushing it to the left or right Remove the Protective Liner from the Sensor Support Mount Place the Sensor vertically on the upper arm 1 2 3 Note Please pay attention to the location of the Sensor site an...

Page 9: ...om the mount Only the Sensor support mount will be left on your body Discard the sensor inserter into a sharp s container or a puncture proof container with a tight lid Connect Transmitter Keep the transmitter parallel to the sensor support mount then snap the transmitter into place The indicator light will flash green 3 times first then another 6 times within one minute 4 5 6 Apply Sensor 8 ...

Page 10: ...ensor Warm up When connected the Sensor will go through a warm up period for 1 hour Start Sensor 9 2 Enter the Sensor code If the Factory Calibration is turned on you can manually enter the 4 digit sensor code on the package of the sensor then the calibrations aren t required If the sensor code is not entered the sensor need to be calibrated twice on the first day ...

Page 11: ...Special Conditions Warm Up The Sensor is warming up ERR The Sensor shall be recalibrated after 15 minutes BG The Sensor shall be calibrated now No readings Do not remove the Sensor until you get Sensor Failure Alert LOST Sensor signal has been lost for more than 10 minutes Wait for a while HIGH Sensor glucose is above 22 2 mmol L 400mg dL LOW Sensor glucose is below 2 2 mmol L 40mg dL Underlined r...

Page 12: ...erted Otherwise wait until the reading recovers SENSOR EXPIRED Change Sensor SENSOR FAILURE Change Sensor METER BG NOW Enter new meter BG for calibration or tap OK to clear the alert SENSOR CAL ERROR Enter meter BG after 15 minutes LOW GLUCOSE Check blood glucose and treat it as necessary Continue to monitor blood glucose LOW PREDICTED Check blood glucose and treat it as necessary Continue to moni...

Page 13: ...DM first Remove the Sensor If the Sensor Expiration alerts are on each sensor session will end automatically after 14 days After that you can remove the sensor from your body Fold and break the sensor support mount and gently pull the Transmitter away from the sensor support mount Remember to keep the reusable Transmitter Make sure the Transmitter was disconnected from the old Sensor over 90 secon...

Page 14: ...se Alerts ON OFF Glucose Limits mmol L Start Time Low Limit High Limit Predictive Alerts Type Status Time High ON OFF Low ON OFF Rate Alerts mmol L min Type Status Rate Rise ON OFF Fall ON OFF Repeat Snooze High Low Others Your Personal CGM Settings 13 ...

Page 15: ...d Shanghai 201203 China Tel 86 21 50274781 Fax 86 21 50274779 This product complies with Directive 93 42 EEC MDD and Directive 2014 53 EU RED MY 158 IM883158WW 001 348502 Version 1 04 Medtrum B V Nijverheidsweg 17 5683 CJ Best The Netherlands Tel 31 0 499745037 EC REP 0197 ...

Reviews: