CHANGING THE MINIMED

™ 

MIO

 ADVANCE INFUSION SET: MiniMed

 630G Insulin Pump

THE BACKLIGHT

MAY HAVE

TURNED OFF

Press any button 

to turn the screen

back on. 

u

5.

With your thumb still on the 

plunger, flip over so vial is on top. 

Release thumb and pull plunger 

down to fill with insulin. 

u

u

u

u

u

Tap the reservoir to move air 

bubbles to top of reservoir. Push 

plunger up to move air into vial. 

u

If needed, pull plunger 

back down to amount 

of insulin needed for 

2-3 days. 

u

To avoid getting insulin on the top 

of the reservoir, turn vial over so it 

is upright. Hold transfer guard and 

turn reservoir counter-clockwise 

and remove from transfer guard. 

6.

7.

8.

2.

1.

Twist plunger 

counter-clockwise to 

loosen and remove. 

4.

Gently push tubing  

connector onto reservoir. 

Turn clockwise until locked.  

You will hear a click. 

3.

If you see air bubbles, tap  

reservoir to move them to   

top. Push plunger just a bit  

to move them into tubing.  

 

Select Next

Select 

Next

.

Insulin 
vial

Plunger

Transfer 

guard

Reservoir

Connector

Plunger

Select 

Load Reservoir

 and 

unlock pump if necessary.

Remove infusion set from package. Remove the 

paper that holds the tubing together and unwind.

u

Continued on next page

u

WARNING:

 

 

Do not use the reservoir or infusion set if any liquid gets on the top of the reservoir or inside the tubing connector. Liquid can 

temporarily block the vents and may result in the delivery of too little or too much insulin, which can cause hyperglycemia or hypoglycemia.

u

CONNECT  

RESERVOIR TO 

INFUSION SET

You will place the 

reservoir connector 

onto the end of the 

infusion set to the 

filled reservoir.

u

Top 
button

Insertion 

device

Raised 

arrows 

indicating 

tubing 

direction

Cannula 

housing

Soft 

cannula

Tubing 

connector

Tubing

Site 

connector

White 

cap for 

site connector

Disconnect 

cover

MINIMED

 MIO

 ADVANCE

 

Infusion Set

Summary of Contents for MiniMed Mio Advance

Page 1: ...SERVOIR CONNECT TO THE INFUSION SET TUBING Follow the next steps to fill reservoir with insulin and connect to the infusion set tubing Remove from package Make sure insulin vial is at room temperature to reduce the risk of air bubbles 1 2 u Pull plunger down to the amount that you plan to fill with insulin 3 Wipe vial with alcohol Place vial on table Firmly press the blue transfer guard onto vial ...

Page 2: ...k 3 If you see air bubbles tap reservoir to move them to top Push plunger just a bit to move them into tubing Select Next Select Next Insulin vial Plunger Transfer guard Reservoir Connector Plunger Select Load Reservoir and unlock pump if necessary Remove infusion set from package Remove the paper that holds the tubing together and unwind u Continued on next page u WARNING Do not use the reservoir...

Page 3: ...hen you see this screen select Next 3 u Select Fill and keep holding until you see drops at the end of tubing then let go Drops at end of tubing 4 After you see drops press and select Next u u INSERTING THE MINIMED MIO ADVANCE INFUSION SET Next follow the steps to insert the infusion set into your body Continued on next page u Remove the paper backing from the adhesive Be careful not to touch the ...

Page 4: ...withyourfinger Replacethe infusionsetiftheadhesivedoes notsticktotheskin u 7 6 Gently hold the cannula housing steady with your finger Then push the site connector straight into the cannula housing until you hear a click u u Gently and carefully remove the insertion device by pulling straight away from your body Dispose of the insertion device in an appropriate sharps container and in accordance w...

Page 5: ... BODY Todisconnect gentlyhold thecannulahousingsteady withyourfinger Then squeezethesidesofthe siteconnectorandpullitout fromthecannulahousing u 2 1 Put the white cap on the site connector u 3 Put the disconnect cover in the cannula housing Then push the cover into the cannula housing until you hear a click 1 Remove the white cap from the site connector by squeezing the side of the site connector ...

Reviews: