2
CHANGING THE MINIMED
™
MIO
™
ADVANCE INFUSION SET: MiniMed
™
630G Insulin Pump
THE BACKLIGHT
MAY HAVE
TURNED OFF
Press any button
to turn the screen
back on.
u
5.
With your thumb still on the
plunger, flip over so vial is on top.
Release thumb and pull plunger
down to fill with insulin.
u
u
u
u
u
Tap the reservoir to move air
bubbles to top of reservoir. Push
plunger up to move air into vial.
u
If needed, pull plunger
back down to amount
of insulin needed for
2-3 days.
u
To avoid getting insulin on the top
of the reservoir, turn vial over so it
is upright. Hold transfer guard and
turn reservoir counter-clockwise
and remove from transfer guard.
6.
7.
8.
2.
1.
Twist plunger
counter-clockwise to
loosen and remove.
4.
Gently push tubing
connector onto reservoir.
Turn clockwise until locked.
You will hear a click.
3.
If you see air bubbles, tap
reservoir to move them to
top. Push plunger just a bit
to move them into tubing.
Select Next.
Select
Next
.
Insulin
vial
Plunger
Transfer
guard
Reservoir
Connector
Plunger
Select
Load Reservoir
and
unlock pump if necessary.
Remove infusion set from package. Remove the
paper that holds the tubing together and unwind.
u
Continued on next page
u
WARNING:
Do not use the reservoir or infusion set if any liquid gets on the top of the reservoir or inside the tubing connector. Liquid can
temporarily block the vents and may result in the delivery of too little or too much insulin, which can cause hyperglycemia or hypoglycemia.
u
CONNECT
RESERVOIR TO
INFUSION SET
You will place the
reservoir connector
onto the end of the
infusion set to the
filled reservoir.
u
Top
button
Insertion
device
Raised
arrows
indicating
tubing
direction
Cannula
housing
Soft
cannula
Tubing
connector
Tubing
Site
connector
White
cap for
site connector
Disconnect
cover
MINIMED
™
MIO
™
ADVANCE
Infusion Set