Seite 18
Operation
Channel B modified V2:
+
white (3)
Left fourth intercostal space at sternal border
-
yellow (2)
Left clavicle at sternal border
Ground (reference potential):
black (5)
right costal arch in the anterior axillary line
Electrode application for patient cable 3, TELESMART-H
Channel A modified V4:
+
green (5)
Left fifth intercostal space in the midclavicular line
-
red (1)
Right clavicle at sternal border
Channel B modified V6:
+
white (6)
Left fifth intercostal space in the anterior axillary line
-
yellow (2)
Left clavicle at sternal border
Channel C modified V2:
+
orange (4)
Left fourth intercostals space at sternal border
-
blue (3)
Left clavicle in the anterior axillary line
Ground (reference potential):
black (7)
Right costal arch in the anterior axillary line
Electrode application for patient cable M, TELESMART
TELESMART-H/P - Holter-ECG-Recording System - version D
Cable 3: red (1), yellow (2), blue (3),
orange (4), green (5), white (6), black
(7)