9
U Manual
•
UVP-I-MULVAC-1
Effective 12-09-16
drying chamber was in use when moving the switch to the
WAND setting, it will be necessary to press START again to
restart the pump. Choose the appropriate vacuum tip which will
allow you to insert it into tube openings, vents and
compartments. Care should be taken so the tubes or
components are not pushed in or dislodged. Press the STOP
switch when finished.
DO NOT OPERATE THE VACUUM WAND WITHOUT THE
FILTER ASSEMBLY.
PRESSURE WAND
The Pressure Wand is active at the same time as the Vacuum
wand. Press START to turn the pump on and press STOP when
finished. If a tip on the vacuum wand is clogged, move it to the
Pressure Wand and turn the pump on. If the vacuum tip
cannot be cleared using the pressure function, turn the unit
off and remove the tip from the wand. A specially designed tool
(marked with red) for cleaning vacuum tips is included in the
enclosed tool kit.
The wand assembly can be removed from the coupling on the
base unit by pushing in the retaining ring and gently pulling on
the tubing. To reinstall, simply insert the tubing back into the coupler.
CARE AND MAINTENANCE
The
U
has been engineered to provide years of
trouble-free service. There are no field serviceable components
inside the cabinet.
Changing the Filter
- The particulate filter has been located on
the Vacuum Wand for ease of service. It should be replaced
when significant amounts of debris are visible. To replace
the filter, grip the end cap of the filter and rotate it to the
left, then pull. Change the filter element and rubber O-ring
with a new one. Replace the end cap and rotate it to the right.
The filter must be in place for the Vacuum Wand to
operate properly.
You
must
use the same type of filter and
O-ring or the performance of the vacuum will be affected.
Should condensation develop in the particulate filter of Vacuum Wand
tube, run the system for 10 minutes to remove.