MedReady 1700PRN Instructions Download Page 2

 

Other Settings and Functions:   

Away Mode 

You may turn the dispenser 

OFF

 by pressing the 

hour

  button and holding for approximately 

4 seconds.  The display will read ‘

OFF’

 

and the medication will stop dispensing.

  

To return to normal 

operation, press the 

hour 

until the PRN time displays.

 

–  

Alarm Test 

Press

 and release this button on the upper left to test the dispensing rotation.  Slide the 

white 

T-BAR

 up reset the PRN time.   

Trouble Shooting 

 

 

No PRN time is being 

displayed, but the colon is 

blinking 

The Device is not getting A/C power.  Try another A/C wall outlet that you 

know is live.   Insure both ends are connected.     

The PRN Time does not 

reset when opening the 

medication door.

 

Instruction # 6 was not followed when replacing the blue lid.   The 

medication door pushes up the white T-BAR (directly below the medication 

door), which triggers a micro switch and resets the PRN time.   If the 

medication door is on the wrong side of the white T-BAR, the medication 

door can’t engage the micro switch to reset the PRN time.

 

The medication tray does 

not rotate when the PRN 

time reaches 0:00

 

The battery is not plugged in.  See step # 1. To test the battery connection, 

disconnect the A/C power, and 

ONLY

 the colon should be blinking.  

 

The new medication slot is 

empty

 

The medication was not loaded properly, Refer to step #5.

 

Display shows ‘OFF’ 

Depress the Hour button by itself for several seconds to return to show the 

PRN time.   

Blue Lid won’t go back on  See # 6 above 
The Battery and A/C are 

connected but the device 

won’t rotate to the next 

medication when the PRN 

time reaches 0:00  

Most likely, the medication tray was previously forced to the next dose 

disabling the device.  Please call our customer support at 310.328.7557 (9-

4:30 M-F Pacific Time) to rectify this situation. 

 

 

Please read and understand the following: 

Note that many PRN regimens involve the dispensing of narcotics. The MedReady 

1700PRN is designed as a medication aid only.  MedReady is NOT a medication Safe or 

Vault. With determination, a patient will be able to break into the device and access more 

than intended.  If there is any doubt that a patient will have an uncontrollable urge to 

access more medication than prescribed, please do NOT use this device.  Under normal 

use, only one dose is available at any one time.

 