MEDRAD
®
Mark 7 Arterion Injection System
15 - 90
18.
Select either
Arm Single
or
Arm Multiple
to arm the injector.
a.
Press any button on the Injector Head.
b.
Ensure the system disarms.
19.
Select either
Arm Single
or
Arm Multiple
to arm the injector.
a.
Use the handle and the back of the Injector Head (but not the Manual Knob) to rotate the
Injector Head to the Purge position (upright).
b.
Ensure the system disarms and cannot be re-armed unless the head is rotated back to
the Injection position.
20.
If a Syringe Heat Maintainer is connected to the Injector Head, ensure that it is warm to the
touch.
a.
Ensure the over-temperature message is not displayed in Sentinel window.
21.
If ISI is connected, enable ISI and check for errors.
22.
Power down the system and remove the syringe and dispose of it properly.
15.3 Annually
Bayer offers Preventative Maintenance Programs. These annual programs greatly assist in maintaining
accuracy and reliability, and can also extend the life of the system. Contact your local Bayer office or
your local authorized dealer for further information. Refer to the back of the title page of this manual for
address, telephone and fax information.
15.3.1 Injection System Calibration
Bayer recommends a complete system calibration and performance checkout be performed annually.
Contact Bayer HealthCare Services, or your local Bayer office for complete details.
15.3.2 Checking Leakage
As part of an annual maintenance program performed by a qualified Services Representative or
authorized dealer, both Electrical leakage and protective earth ground continuity checks should be
performed.
NOTE:
Local regulations or hospital protocol may require electrical leakage checks at
more frequent intervals. If this applies, the local regulations for leakage must be
followed.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......