MEDRAD
®
Mark 7 Arterion Injection System
10 - 58
6.
Stop aspirating with manual knob once blood is visualized in the high pressure connector
tubing.
7.
Verify that there no air is in the high pressure connector tubing.
a.
If there is air in the tubing, disconnect from patient, remove air, and try fluid to fluid
connection again.
8.
Once it has been established that all air has been removed, turn the Manual Knob clockwise
to clear blood from high pressure connector tubing.
The system is ready to define protocols or arm.
10.7 Enabling 15 mL Purge Feature and Choosing Configuration Options
The operator can choose between two different configurations (
ON
or
OFF
) of the 15 mL Purge Option.
NOTE:
Access the 15 mL Purge Configurations (
ON
or
OFF
) from the
Options
menu.
NOTE:
Default configuration upon power-up is 15 mL Purge
ON
.
10.7.1 15 mL Purge ON
•
The system will require a purge of 3.5 mL or more with the Injector in the upright position
with a plunger retraction of 15 mL or more, regardless of whether the Injector head is in the
upright or downward position.
•
The system will
NOT
require a purge in the upright position with a plunger retraction of less
than 15 mL, regardless of whether the Injector head is in the upright or downward position.
•
Auto-Fill
will remain available for use with 15 mL Purge turned
ON
.
•
With 15 mL Purge turned
ON
, the system will require a purge in the upright position during
system power-up, syringe installation, and new case.
NOTE:
The operator can still choose to purge prior to arming, if desired, with a plunger
retraction of less than 15 mL but the system will not require it.
10.7.2 15 mL Purge OFF
•
The system will require a purge of 3.5 mL or more in the Injector’s upright position
ONLY
with
a plunger retraction of 15 mL or more in the Injector’s upright position.
•
Auto-Fill
will be available for use
ONLY
in the Injector’s upright position.
•
With 15 mL Purge turned
OFF
, the system will require a purge in the upright position during
system power-up, syringe installation, and new case.
NOTE:
The operator can still choose to purge, if desired, with a plunger retraction of less
than 15 mL and/or upon plunger retraction in the Injector’s downward position.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......