Setting and Managing Protocols
9 - 41
9
Setting and Managing Protocols
This chapter discusses how to:
•
"Set Injection Parameters from the Home Tab"
•
"Manage Protocols from the Protocols Tab"
9.1 Set Injection Parameters from the Home Tab
NOTE:
Refer to syringe and disposable packaging to confirm lowest pressure rating prior
to programing an injection.
9.1.1 Set Injection Parameters on Home Tab - Single
A Single protocol injects a set Volume of contrast at one flow rate, pressure, rise time, and delay time
(if using ISI).
NOTE:
Delay and rise time are not available for mL/m protocols.
1.
Select the
Single
tab (A).
Figure 9 - 1: Set Single Protocol
Over Volume Hazard - Serious patient injury or death may result.
•
Do not program a protocol outside of the clinically accepted volume range.
•
Ensure that the correct volume is programmed in the protocol for the target anatomy.
Vessel Dissection Hazard - Serious patient injury or death may result.
•
Do not program a protocol outside of the clinically accepted flow rate range.
•
Ensure that the correct flow rate is programmed in the protocol for the target anatomy.
•
Do not program a protocol outside of the clinically accepted pressure limit.
•
Ensure that the correct pressure limit is programmed in the protocol for the target
anatomy.
Foreign Body Embolism Hazard - Serious patient injury or death may result.
•
Do not program a pressure greater than the lowest rated disposable’s pressure rating.
Bloodborne contamination hazard - Serious patient and/or worker injury or death may
result.
•
Do not program a pressure greater than the lowest rated disposable’s pressure rating.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......