Installation - System and Accessory
16 - 103
16.5.4 Wall Mount Bracket Installation
NOTE:
Before attaching the bracket to the wall, the installer must know if the wall studs in
the room are made of metal or wood. This will determine the kind of hardware
required to complete the installation.
NOTE:
The bracket must be mounted to a stud to insure a secure mount. Wallboard and
wing anchors will not support the weight and movement created by swinging the
control console into position during routine use.
NOTE:
Do not install the wall mount bracket on a wall which contains shielding unless the
installation does not violate the shield.
NOTE:
If the wallboard is thicker than 5/8 inch, longer screws must be purchased to
insure a secure mount.
1.
Hold the bracket with the friction plates on the bottom and position the wall bracket on the
wall where it is to be installed and mark the mounting holes. The bracket must be mounted
to a stud to insure a secure mount.
2.
Drill a 5/32” by 2” deep hole through the wall and into the stud (3/8” hole if there is a metal
stud).
3.
On metal studs only, place the hollow wall anchors into the holes drilled.
4.
Place the wall bracket onto the wall and secure with the fasteners provided. (#12 x 2” screws
for wood studs, #10-24 x 2 1/2“screws for metal studs)
5.
Remove the four screws from the back of the Display Control Unit.
Figure 16 - 23: Remove Screws
6.
Leave the cable retainers (A) attached to their respective cables.
7.
Place the Display Control Unit onto the plate and secure the Display Control Unit to the
bracket with the #8-32 x 5/8” screws provided, as shown in Figure 16 - 24. Ensure to attach
the cable retainer clips to the upper right and lower right holes on the Display Control Unit.
Figure 16 - 24: Wall Mount Screws and Mounted Display Control Unit
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......