D) To reset the pager slide the pager side switch to “off/reset”. If the Medpage monitor
alarm is active do not switch the pager(s) on until the monitor alarm has been reset.
SETUP MODE
I M PORTANT NOTE: THI S PROCEDURE I S ONLY NEEDED
FOR SETUP, NOT FOR NORM AL OPERATI ON
General Note: The alarm state is indicated when both the
POWER and M OVEM ENT LEDs are displaying solid RED.
When the LEDs are no longer both RED, the alarm has reset.
When the monitor is first powered up, it may not respond to
sound or movement as the internal calibration period can
take 10 – 15 seconds. After this time, the monitor will begin
detection.
The main monitor can be reset at any time during this setup and you can safely start again
should you not feel comfortable with anything. To turn the monitor off push the on/off button
and wait 5 – 10 seconds until the POWER and MOVEMENT LEDs turn off, the monitor is
now safely switched off.
In order to enter setup mode take the following steps, please read through all 7 steps before
beginning;
1
. Turn the delay control to maximum
2
. Make sure the monitor is turned off
3
. Whilst performing this step, please watch the
M OVEM ENT
and
POWER
LEDs. Now
hold in the red setup button on the back of the monitor, and then turn the unit on whilst still
holding the setup button, keeping the setup button held in.
4
. After a short period the LED’s on the front panel will begin to alternately flash
RED/ORANGE, you can now release the setup button
5
. The monitor is now in setup mode, you can now adjust the settings to suit your
requirements.
6
. The next stage is to decide how you would like the monitor to alarm. The MP5x can
monitor movements as well as sound to raise an alarm. In order to reduce false alarms a delay
can be set, this will prevent the alarm being raised too soon. Below is an example:
If a person rolls over during the night, this creates a lot of movement on the sensor. An
MP5x User Instruction REV 01-12-17