medotech GRINDCARE 3.0 User Manual Download Page 11

11

Contraindications

Patients with a pacemaker are not allowed to use GRINDCARE.

Warnings

  1.

  The long-term effects of chronic use of GRINDCARE are unknown.

  2.

   The GRINDCARE electrode is only intended to be placed over the temporalis 

muscle on the temple.  The electrode should not be applied in the following 
areas: 
a.  over the carotid sinus nerves (on the side of the throat),

   

b.   on the chest since the introduction of electrical current into the heart may 

cause cardiac arrhythmias,

   

c.  over  the cerebrum ( brain),

   

d.   over swollen, infected or inflamed areas or skin eruptions, e.g. phlebitis,  

varicose veins, etc.

   

e.  over or near cancer lesions.

  3.

   Simultaneous connection of a patient to high-frequency surgical equipment 

may result in burns at the site of the Stimulator electrodes and possible  
damage to the Stimulator.

  4.   Operation in close proximity (e.g. 1 m, 3 ft.) to short wave or microwave  

therapy equipment may produce instability in the Stimulator output.

 
The following precautions must be observed when using  
GRINDCARE:

When using GRINDCARE together with a hearing aid, the function of the hearing 
aid may be impaired.

Approximately 5% of the population will have an allergic reaction to electrode  
gel, i.e. redness, rash or irritation at the point of contact. In such case, remove  
electrode, wash off any gel residue and contact your clinician.

Electric blankets and magnetic bed linen must not be used when the device is  
in use. 

The GRINDCARE warranty will be void in the event of failure to 
comply with the following:

 

1.

  The device must not be subjected to jolts, knocks or other mechanical impacts.  

 

2.

    The electric unit must not be immersed in any fluid and must be protected 

against moisture.

 

3.   

The device must be cleaned in accordance with the instructions. See the 
section on maintenance. 

  4.

   The performance of the device may be impaired if it is exposed to 

temperatures below 10°C (50°F) and above 40°C (104°F).

  5.

  The device must not be dismantled. 

  6.

  Only the charger supplied may be used to charge the device. 

  7.

   Only the supplied battery may be recharged. If it needs replacing, 

return the device to the supplier. 

 

8.   

During transportation, the device must be kept in the original packaging 
supplied, otherwise the warranty will be void. 

  9.

   If the device stops working as intended for any reason, the supplier must 

be contacted. 

 

Safety meaSureS

Summary of Contents for GRINDCARE 3.0

Page 1: ...User Manual GRINDCARE 3 0 US...

Page 2: ...2...

Page 3: ...3 introduction 4 GENERAL INSTRUCTIONS 5 usage 6 DATA 7 CONFIGURATION 8 trouble shooting 9 Maintenance 10 TECHNICAL SPECIFICATIONS 13 Contents...

Page 4: ...which is connected to the Stimulator and placed on the temporalis muscle on the temple This electrode measures electrical signals from the muscle when it moves and emits an electrical impulse which ma...

Page 5: ...rode is now correctly placed and you can start calibrating GRINDCARE see page 6 8 Before use GRINDCARE must measure how much you grind your teeth Find out more about this baseline function in the sect...

Page 6: ...n has now started and there is no wireless communication between the Stimulator and Grinddock You can place Grinddock in its stand it will switch to standby after 5 sec Next morning Carefully remove t...

Page 7: ...o view the results of your last activation according to How much you have been grinding your teeth in total Grinds How many hours your activation lasted Hours How many times an hour you have been grin...

Page 8: ...tings by pressing OK Choose your required language using the arrow buttons Approve your settings by pressing OK configuration Select Clear Data if you want to delete all activation data In Set Sensiti...

Page 9: ...nals maintain good contact with the skin Review your activation settings again and try to bite down harder This allows you to set a higher threshold for when GRINDCARE is sending bio feedback 3 I find...

Page 10: ...thanol propylalcohol or similar Gel pads These are for single use only and must not be cleaned These must be replaced every night before use of GRINDCARE Storage of gel pads The properties of the gel...

Page 11: ...the function of the hearing aid may be impaired Approximately 5 of the population will have an allergic reaction to electrode gel i e redness rash or irritation at the point of contact In such case re...

Page 12: ...ers or hearing aids See the section on contraindications Type BF equipment IP classification CE labelling applicable in accordance with directive MDD 93 42 EEC The equipment must not be exposed to moi...

Page 13: ...0 days activation data Charger 100 240 V AC 50 60 Hz Battery 3 7 V Lithium Polymer battery Average battery charge time 3 5 hours Anticipated operating time between charges 5 days Temperature during op...

Page 14: ...14...

Page 15: ...15...

Page 16: ...Medotech s core expertise relates to the development of custom devices capable of recording and analysing signals from the human body GC3 User Manual US Complaints Complaints may be submitted in acco...

Reviews: