Page 2: ...2...
Page 3: ...3 introduction 4 GENERAL INSTRUCTIONS 5 usage 6 DATA 7 CONFIGURATION 8 trouble shooting 9 Maintenance 10 TECHNICAL SPECIFICATIONS 13 Contents...
Page 4: ...which is connected to the Stimulator and placed on the temporalis muscle on the temple This electrode measures electrical signals from the muscle when it moves and emits an electrical impulse which ma...
Page 5: ...rode is now correctly placed and you can start calibrating GRINDCARE see page 6 8 Before use GRINDCARE must measure how much you grind your teeth Find out more about this baseline function in the sect...
Page 6: ...n has now started and there is no wireless communication between the Stimulator and Grinddock You can place Grinddock in its stand it will switch to standby after 5 sec Next morning Carefully remove t...
Page 7: ...o view the results of your last activation according to How much you have been grinding your teeth in total Grinds How many hours your activation lasted Hours How many times an hour you have been grin...
Page 8: ...tings by pressing OK Choose your required language using the arrow buttons Approve your settings by pressing OK configuration Select Clear Data if you want to delete all activation data In Set Sensiti...
Page 9: ...nals maintain good contact with the skin Review your activation settings again and try to bite down harder This allows you to set a higher threshold for when GRINDCARE is sending bio feedback 3 I find...
Page 10: ...thanol propylalcohol or similar Gel pads These are for single use only and must not be cleaned These must be replaced every night before use of GRINDCARE Storage of gel pads The properties of the gel...
Page 11: ...the function of the hearing aid may be impaired Approximately 5 of the population will have an allergic reaction to electrode gel i e redness rash or irritation at the point of contact In such case re...
Page 12: ...ers or hearing aids See the section on contraindications Type BF equipment IP classification CE labelling applicable in accordance with directive MDD 93 42 EEC The equipment must not be exposed to moi...
Page 13: ...0 days activation data Charger 100 240 V AC 50 60 Hz Battery 3 7 V Lithium Polymer battery Average battery charge time 3 5 hours Anticipated operating time between charges 5 days Temperature during op...