-29-
Guidance and manufacturer’s declaration - electromagnetic immunity
The [MDS3003]/[MDS3003P] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [MDS3003]/[MDS3003P] should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601test
level
Compliance
level
Electromagnetic environment -
guidance
Radiated RF
IEC 61000-
4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the [MDS3003]/[MDS3003P], including
cables, than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.5 GHz