background image

Pulse Oximeter 

 

   

General Description                                                                 

The measurement of oxygen saturation of arterial blood (also known 

as  pulse  oxygen  saturation,  usually  shortened  as  SpO

2

)  adopts  the 

principles of light spectra and volume tracing. The LED emits lights with 
two specific wavelengths, which are selectively absorbed by oxygenated 
hemoglobin  and  deoxyhemoglobin.  The  optical  receptor  measures  the 
changes in the light intensity after the light passes the capillary network 
and  estimates  the  ratio  of  oxygenated  hemoglobin  and  the  total 
hemoglobin. 

 

 
 
 

 

 

Precautions for use                                                                   

!

  Explosion hazard. Do not use the oximeter in the presence of 

flammable anesthetics mixture with air, oxygen, or hydrogen. 

!

  When  the  oximeter  is  in  use,  there  should  not  be  any  great  power 

appliances  as  high  voltage  cables,  X-ray  machine,  ultrasound 
equipment and electrizer in use nearby.   

!

  Keep  the  oximeter  away  from  dust,  vibration,  corrosive  substances, 

explosive materials, high temperature and moisture. 

!

  This oximeter does not have alarm function; please do not use this 

product in the environment where alarm is required. 

!

  The oximeter should be handled with care so as to avoid shocks and 

falls.   

!

  When the oximeter is in use, it must be ensured the batteries have 

sufficient capacity; otherwise there might be such phenomena as 
starting-up abnormalities or inaccurate measurement data, etc.     

!

  Please  do  not  use  such  pointed  objects  as  pen  point  or  nails  for 

pressing operation, otherwise it might cause permanent damage to the 
surface of the keyboard. 

!

  Do not make any clinical judgments based solely on the oximeter. The 

oximeter is intended only as an adjunct in patient assessment. It must 
be used in conjunction with clinical signs and symptoms, as well as 
doctor’s diagnoses. 

!

  To ensure accurate performance and prevent device failure, do not 

expose the oximeter to extreme moisture, such as direct exposure to 
rain. Such exposure may cause inaccurate performance or device 
failure. 

!

  Do  not  conduct  SpO

2

  measurement  on  the  finger  smeared  with  nail 

polish, otherwise this will lead to unreliable measurement results. 

!

  Please do not open the enclosure. The enclosure shall only be opened 

by the authorized person. 

!

  In  order  to  have  more  accurate  measurements  of  SpO

2

  and  PR,  the 

oximeter should be used in quiet and comfortable environment. 

!

  Follow local ordinances and recycling instructions regarding disposal or 

recycling of the device and device components, including batteries. 

 

   

Intended Use                                                                                     

The  Fingertip  Pulse  Oximeter  is  intended  to  measure  functional 

arterial  oxygen  saturation  (SpO2)  and  pulse  rate  of  adult  patients  in 
hospital, hospital type facilities, as well as in the home care environment. 

The oximeter is not suitable to monitor patient continuously for long 

term

 

 

 

 

Operation Instructions                                                                     

1. 

Install  two  AAA  batteries  into  battery  cassette  before  closing  its 
cover. 

2. 

Nip the oximeter, then insert one of fingers   

into the rubber hole of the oximeter before   
releasing the oximeter, and your nail surface

 

must be upward.

                                             

 

3. 

Press the function button once on front panel. 

4. 

Your finger and body do not tremble during measuring.   

5. 

Read corresponding data on the display screen. 

6. 

After turning on the oximeter, each time you press the power switch, 
the display screen will change to another direction. 

 

 

SpO

2

 Plethysmogram 

 

SpO

reading 

 

Pulse rate reading 

 

Indication of battery capacity 

 

Indication of pulse intensity 

 

    Battery Installations   

                                                                 

1.

 Press the button down on the back panel of oximeter (for M70, you can 

just skip this step to move on the next). 

2.

 Push the battery cover horizontally along the arrow shown on the back 

panel of oximeter. 

3.

 Install the two AAA batteries into battery cabin   

in correct polarities.   

4.

 Close the battery cover.   

 
 
 

 

Notes: 

!

 

Please put or remove batteries in right order, or it is likely to 
damage the device bracket. 

!

 

Battery polarities must be correctly installed. Otherwise, damage 
might be caused to device. 

!

 

Please remove the battery if the oximeter will not be used for long 
time. 

 

   

Maintenance                                                                                   

1.

  Use  a  soft  cloth  dampened  with  either  a  commercial,  nonabrasive 

cleaner, or a solution of 70% alcohol in water, lightly wipe the surfaces 
of the oximeter. 

2.  The  most  commonly  used  hospital  cleaning  agent  and  non-corrosive 

detergents  can  be  used  for  cleaning  the  oximeter,  but  please  be 
careful  that  many  types  of  detergents  must  be  diluted  before  use; 
Please  use  them  according  to  the  directions  of  the  manufacturers  of 
the detergents.   

3. Avoid using alcohol-based, amido or acetone-based detergents   
4.  The  casing  of  the  oximeter  should  be  kept  from  the  contamination  of 

filth  and  dirt,  and  it  can  be  wiped  with  non-velvet  soft  cloth.  When 
cleaning, do not spill the liquid onto the instrument. Ensure no liquid is 
allowed to enter the inside of the oximeter.     

5.  It  is  forbidden  to  use  such  grinding  materials  as  wire  brush  or  metal 

polishing  agent,  because  these  materials  may  cause  damage  to  the 
panels of the oximeter.   

6. Please do not soak the oximeter in liquid. 
7. Under normal circumstances, it is unnecessary for the oximeter to have 

special maintenance, and cautions must be exercised on the following 
points during the use of the oximeter:   

!

  Please  use  the  oximeter  in  the  environment  according  to  the 

requirements of the performance criteria. 

!

  Avoid exposure or direct sunlight. 

!

  Avoid excessive radioactive infrared rays or ultraviolet rays. 

!

  Avoid contacts with organic solutions, dusts or corrosive gases.

 

 

   

Product Specifications                                                                 

"

 

Measurement specifications 

SpO

2

 

Measuring Range 

0~100% 

Resolution 

1% 

Accuracy 

At 70~100%, ±2%

 

At 40~69%,±3%

 

At 0~39%, unspecified   

Data update period 

<13 s 

PR 

Measuring Range 

25~250 bpm 

Resolution 

1 bpm 

Accuracy 

±1% or ± 1 bpm, whichever is greater 

Data update period 

< 13 s 

"

 

Battery specifications 

Type 

Voltage 

two AAA alkaline battery 

1.5 Volts DC (per battery) 

The oximeter uses two 1.5 V A AA type batteries and a set of new 

 

 

 

 

 

%

100

lobin

deoxyhemog

 

bin 

oxyhemoglo

hemoglobin

 

oxygenated

%

SpO

2

×

+

=

Reviews: