TROUBLESHOOTING

Symptom

Possible 

Cause

Remedy/Solution

None of the func-

tions work

Hand pendant?

Is hand pendant 

plugged in?

Plug in hand pendant

Did you try another 

pendant from working 

bed

Use different hand       

pendant from a  

working bed

Does the bed 

have staff 

control?

Are the lock-out    

functions engaged?

Disengage lock-out 

functions

Is the staff control 

wired to the control 

box properly?

Plug hand pendent 

directly into “HB” 

port of control box

Power?

Is the bed plugged into 

an outlet?

Plug bed into wall outlet

Is the power outlet on 

a switch?

Turn switch on, check fuse 

box, or try another wall 

outlet

Verify bed power source 

(outlet) is adequate for 

bed electrical rating

Has the power cord 

become loose from the 

control box

Insert power cord fully 

into control box

Is the power cord 

damaged?

Replace power cord

Functions work but 

bed moves slowly

Power?

Is the bed plugged into 

an outlet?

Plug bed Into wall outlet

Is the power outlet on 

a switch?

Turn switch on, check fuse 

box, or try another wall 

outlet

Verify bed power source 

(outlet) is adequate for 

bed electrical rating

Has the power cord 

become loose from the 

control box?

Insert power cord fully 

into control box

Is the power cord 

damaged?

Replace power cord

Grinding noise 

heard from one of 

the actuators

Actuator gear is 

stripped

Actuator gear is 

stripped, replace 

actuator

Replace actuator

TROUBLESHOOTING (CONTINUED)

Symptom

Possible 

Cause

Remedy/Solution

One of the actuators 

is not functioning

Hand pendant?

Try another pendant 

from working bed

Use a different hand 

pendant from a working 

bed

Actuator?

Insert actuator cord 

into another working 

port of the control box

If bed still isn’t functioning, 

replacing actuator 

Actuator cable? Verify actuator cable is 

not damaged

If damaged, or Cut replace 

actuator cable

Verify actuator cable 

plug is inserted 

completely in actuator 

and control box

Push actuator cable plug 

firmly into actuator/con-

trol box

Control Box

Insert  actuator cord 

into another working 

port on the control box

If actuator works on a 

different port of the

control box

Bed is not level 

when operating 

High-Low functions

Bed load is not evenly 

distributed

Raise or lower bed to its 

maximum or minimum 

height

Actuators operate 

but wrong move-

ment occurs

Control box 

wiring?

Actuators are plugged 

into the wrong ports

Plug actuators into proper 

ports on control box

Replace control box

IF USE OF THIS TROUBLE SHOOTING GUIDE DOES NOT CORRECT THE PROBLEM, PLEASE CALL 
MEDLINE TECHNICAL SUPPORT AT 1-800-MEDLINE.

  CAUTION 

 

SERVICE / MAINTENANCE PERSONNEL 
The bed is a heavy piece of equipment. It is strongly recommended that any servicing or 
maintenance of the bed be performed by two or more personnel.

Summary of Contents for ALTERRA 1385

Page 1: ...CE MANUAL ALTERRA 1385 LONG TERM CARE BED FCE1385B This product is not made with natural rubber latex Packaging Handling Instructions see Page 8 Shown with optional side rails mattress head footboards...

Page 2: ...cific potential entrapment zones in hospital bed systems Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment The entrapment zones represent the relationship of the various bed...

Page 3: ...IS BED WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT YOUR MEDLINE SALES REP OR MEDLINE CUSTOMER SERVICE BEFORE USIN...

Page 4: ...r transferring in or out of bed DO NOT ALTER THE PLUG ON THE ELECTRICAL CORD If the plug does not fit the existing outlet a correct outlet should be installed by a licensed electrician NEVER operate t...

Page 5: ...into the head section H frame using a rubber mallet if necessary INSTALLING OPTIONAL HEAD AND FOOTBOARDS Tools Required Philips Screw Driver Rubber Mallet Materials Required Head and Footboards and H...

Page 6: ...nt 2 C Clips 2 Locate and install C clips into grooves on the rods of the head section assist rails until they are fully engaged Place C clips on each side of the deck section rod tab as shown below 3...

Page 7: ...in bodily injury The battery back up is a factory sealed unit DO NOT ATTEMPT to open repair unit Contains sealed lead battery Battery must be recycled REPLACING BATTERY Tools Required Flat Head Screw...

Page 8: ...place power cord Grinding noise heard from one of the actuators Actuator gear is stripped Actuator gear is stripped replace actuator Replace actuator TROUBLESHOOTING CONTINUED Symptom Possible Cause R...

Page 9: ...work platform 1 Remove all accessories mounted to the bed assist rails assist bars and head and footboards 2 Elevate the bed to its highest position and raise the head and foot sections 3 Unplug the...

Page 10: ...Section Bed Deck 17 FCE1385017 1 Wall Guard 18 FCE1385018 1 Head Section H Frame 19 FCE1385019 1 Foot Section H Frame 20 FCE1385020 1 Thigh Section Bed Deck 21 FCE1385021 2 Leg Assembly 22 FCE1385022...

Page 11: ...y examine all welds for cracks 3 Inspect the deck components for any bending damage or excessive wear 4 Check the ends of the actuators and mounting hardware for any bending damage or excessive wear 5...

Page 12: ......

Page 13: ...ir Installation and freight costs for replaced or repaired items remain the responsibility of the Customer This warranty does not cover failure due to negligence accident deliberate abuse nor improper...

Reviews: