Troubleshooting Guide
Situation
Possible Cause
e
Solution
Battery light is on
Pump is operating on battery power
Plug cord into AC outlet
LOW BAT alarm activated
Battery charge is below performance level
Plug cord into AC outlet
LOW BAT light is on
Battery life is fifteen minutes or less
Plug cord into AC outlet
To avoid degrading the battery life keep battery charged
If low battery persists then return pump for service
OCCL/EMPTY
alarm activated
Feeding container is empty
Replace or refill feeding container
Occlusion is restricting flow
Identify point of occlusion
- Pump set
•kinked tubing
•blocked tubing
•closed roller clamp
- Feeding tube
- Feeding container
Set drip chamber not properly placed in pump
Assemble properly
Pump is not in an upright/vertical - position or not clamped
properly to the pole
Straighten the pump and clamp properly to the pole
Pump is on hold for more than 2½ minutes**
Press RUN/HOLD button
Bright light is reaching drop sensor
Turn pump so sensor is not in direct light
Drip chamber walls are coated with formula or condensation
Manipulate chamber to clear drops in sensor path
Drop sensor lenses in drip chamber guide are dirty
Clean with cotton swab dampened with isopropyl alcohol
Drip chamber is overfilled
Replace with new Administration Set