11
Chapter6 Notes Before Operation
6.1 To use EKG312T safely and effectively, you should read the user manual carefully before
operation.
6.2 Notes for installation and storage:
1)
EKG312T shall be keep away from high voltage cable, X-ray equipment, ultrasound device
or electrotherapeutics machine,etc.
2)
Avoid using and storing EKG312T in the places with high air pressure, temperature and
humidity beyond standard, bad ventilation, excessive dust or salt-alkali gas and chemical
medicine.
6.3 EKG312T should be put on flat place and taken and put lightly when moved. Avoid too strong
vibration and shock.
6.4 AC frequency and voltage value should be in accordance with the requirements to ensure
enough current capacity.
6.5 Please place EKG312T indoor where it is easy to be grounded. Do not connect the patient and
the cables with other conductors including ground or beds which can be conducted well with
ground.
6.6 Clean the lead cables with water and soap, and sterilize with ethanol and aldehyde base.