19
E.g. C-SCANL means Digital Portable Diagnostic Ultrasound System with L-linear array
2.1.2
Performance
Device specifications
Volume:165.3 x 85.4 x 19.8 mm
Probe length:1150 mm
Weight:L:550g/C:570g
Display size:6 inch LCD
Power specifications
Power supply:DC3.8V,2500mA
Lithium battery capacity:5400mAh
Charging conditions:AC/DC power adapter,input:AC100V-240V,
50/60Hz;output:DC5V,2000mA
B model Performance:
B mode
Probe
information
Probe Type
Convex array(type:C5-2) Linear array(type:L10-5)
Radius of
curvature
60mm
--
Probe center
frequency
MHz
3.5
7.5
Image
performance
indicators
Sound
operating
frequency
The deviation of the sound
operating frequency from
the nominal center
frequency shall be within ±
10%, i.e. the frequency
range is from 3.15 to 3.85
MHz
The deviation of the sound
operating frequency from
the nominal center
frequency shall be within ±
10%, i.e. the frequency
range is from 6.75 to 8.25
MHz