30
Issue
Possible Cause
Solution
No power or
display is blank.
The power supply
is not connected.
Power is not
turned on.
Plug the power supply
into a wall outlet.
Ensure the cable is
plugged into the
Controller Unit. The
Unit will beep once the
power is connected.
Press the Power button.
System beeps
and displays
an
E01
error.
The Treatment Array
is not connected
properly.
Unplug and re-plug the
Treatment Array and press
the start/pause button
on the Controller Unit to
continue the treatment.
If problem persists,
contact BioFlex Service.
System beeps
and displays
an
E02
error.
The Treatment Array
is hot (temperature
exceeds 48°C) due
to being covered or
running too long.
In the event that the
Treatment Array gets too
hot, it will shut off until the
temperature has dropped.
The Treatment Array
needs to be cooled
for at least 15 minutes
before it is used again.
System beeps
and displays
an
E03
error.
The Teatment Array
cable may have
been removed
unexpectedly.
The temperature
sensor has
malfunctioned.
Make sure the cable
is properly connected
to the Controller.
If problem persists,
contact BioFlex Service.
System beeps
and displays
an
E04, E05,
E06, E07, E08,
E09, E10, E11
or E12
error.
The system has
malfunctioned.
Record the error code and
contact BioFlex Service
for further assistance.