52
Manual and Manufacturer's Declaration – Electromagnetic Emission
Emission test
Conformity
High-frequency emission
CISPR 11
Group 1
Class B
Harmonic radiation
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emission
IEC61000-3-3
Complies
Manual and Manufacturer's Declaration – Electromagnetic Immunity
Emission test
Conformity
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV for contact discharge
± 15kV for air discharge
Radiated, radio-frequency,
electromagnetic field
IEC 61000-4-3
Proximity field from RF wireless
communication equipment IEC 61000-
4-3
3 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
See Table 1
Electrical fast transient / pulse group
IEC 61000-4-4
± 2 kV for mains supply line
repetition frequency 100 kHz
Shock pulse
IEC 61000-4-5
± 1 kV line – line
± 2 kV line – earth
Immunity to conducted disturbances,
induced by RF fields
IEC 61000-4-6
3 V (0.15 MHz – 80 MHz)
6 V in the ISM bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
Mains frequency magnetic field (50/60
Hz)
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions and
voltage variations on the mains supply
line
IEC 61000-4-11
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0% UT; 1 cycle and 70% UT; 25/30
cycles
1 phase: at 0 °
0% UT; 250/300 cycles
Summary of Contents for M301V01
Page 1: ...Rev A 25 6 2021 VIVERA Monoblock M301 M301V01 Operation and Maintenance Manual...
Page 12: ...12 Modification M01 Modification M04 0 50 30 30 0 30 0 90 0 30 0 90...
Page 13: ...13 Modification M05 Modification M08 0 50 30 30 0 30 0 90 0 30 0 90...
Page 54: ...54 Maintenance Records Date Description of maintenance performed Performed by...