Cleaning and Maintenance
Warning:
In order to avoid fires, oxygen must not come into contact with oil or grease. It is therefore essential to ensure
that all parts of the medinCNO
®
and the CPAP system which come into contact with oxygen (e.g. gas feed tubes,
O
2
inlet to the medinCNO
®
, patient outlet, patient tubes and Medijet
®
) remain free from oil and grease even
during cleaning, repairs and maintenance.
Medin offers two difference warranty extensions:
-
3 years (REF 39-912) from the date of production
-
5 years (REF 39-913) from the date of production
These warranty expansions only relate to components with defects caused by production, wear parts, as well
as software problems. In addition, the customer must demonstrate that all required maintenance was
performed.
Please note:
The following are not included in the warranty: physical damage, environmental damage (e.g. water or grease
in the device connection, excessively high voltage), improper use as well as consumables (e.g. oxygen cells,
filters, batteries).
9.1
Cleaning
The surface of the medinCNO® should be disinfected before first use using a slightly damp cloth
and every
time after it has been used on a patient. 70% isopropyl alcohol solution can be used for this purpose.
Warning:
-
The device must never be sterilised or immersed in a liquid solution.
-
No abrasive substances may be applied to the surface.
-
No liquid may penetrate the device.
9.2
Maintenance and exchange of the internal batteries
The medinCNO
®
must undergo maintenance and a function check in accordance with the manufacturer's
instructions every 12 months. This must be carried out by trained specialists using appropriate measuring and
testing equipment.
The following actions must be performed during this maintenance:
-
The oxygen cells are regenerated or their service life is checked.
o
Type OOM-102 (REF OOM-102) oxygen cells must be replaced annually.
o
Type MLF-16 (REF MLF-16) oxygen cells must be replaced if they have less than one
year of service life left or no later than after three years.
-
Test of the device’s functionality and sensor calibration.
1
Slightly damp = Take a wet cloth and wring it out. Then wrap a dry cloth around the wet cloth and wring out
both of these cloths together once again.