+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
11
von
68
3.2
Power supply
3.2.1
Mains operation
The power supply unit (REF 39-113) of the medinCNO
®
may only be connected to a power supply with the
following properties:
•
Voltage: 100 – 240 V~ (alternating current)
•
Frequency: 50 – 60 Hz
•
Strength of current: at least 1.1 A
The power supply unit of the medinCNO
®
has a C8 connector for connection to the mains grid. The power line
between the power supply unit of the medinCNO
®
and the mains grid must therefore meet these conditions:
•
Power supply side: plug appropriate for the country
•
Device side: C7 plug in accordance with EN60320
•
Minimum rated voltage: the minimum rated voltage of the power line must be at least as high as
the voltage of the local mains supply. (Recommended: 240 V)
•
Minimum rated current: 1.1 A
•
Maximum length 2 metres (in order to achieve EMC in accordance with DIN EN 60601-1-2:2016)
Spare power lines with EU plugs can be ordered under REF 39-115 (supplied as standard as part of the power
supply unit). Alternative power lines for other countries are available under these numbers:
•
with UK plug
REF 39-116
•
with plug for USA/Japan
REF 39-117
•
with Australian plug
REF 39-118
•
With Chinese plug
REF 39-127
Attention:
The medinCNO
®
may only be run on the local mains grid using the power supply unit provided by the
manufacturer (REF 39-113) and suitable power lines, otherwise the functionality and EMC of the device cannot
be guaranteed.
3.2.2
Battery operation
The medinCNO
®
contains an internal battery and can work on a fully charged battery for up to three hours.
Summary of Contents for medinCNO