background image

SIDE

   

      Display Messages And Problem-Solving Guide

 9

.

     Control Solution Testing

If your Medicare control solution did not come with an Medicare control solution 

Instruction, the following information shall take the place of the instruction.

 

Medicare control solutions contain a pre-set amount of glucose that reacts with 

OKmeter test strips. By testing your control solution and comparing the test results 

with the expected range printed on the test strip vial label, you can make sure that the 

meter and the test strips are working properly together as a system and that you are 

performing the test correctly. 

Why perform a control solution test?

1.  To ensure that your meter and test strip are working properly together.

2.  To allow you to practice testing without using your own blood.

When should the control solution test be performed?

1.  Whenever you suspect that the meter or test strips are not working properly.

2.  When your blood glucose test results are not consistent with how you feel, or 

when you think your results are not accurate.

3.  When test strips are exposed to extreme environmental conditions.

4.  If you drop the meter.

5.  After changing the battery. 

  Important Control Solution Information

1.  Check the expiration date on the control solution bottle. Do not use if expired.

2.  Control solution, meter, and test strips should come to room temperature (68-

77 ºF/20-25ºC) before testing.

3.  Shake the bottle before use, discard the first drop of control solution after 

squeezing, wipes off the dispenser tip to avoid contaminations. These steps 

ensure you will get a good sample and an accurate result.

4.  Record the discard date on the bottle when you open a new bottle of control 

solution.

NOTE :

  

1. There are two levels of control solution (normal and high) available to purchase.  

Please contact with your local distributor when required.

                2. The control solution range printed on the test strip vial is for Medicare control solution 

only. It is used to test meter and strip performance. It is not recommended range for 

your blood glucose level. 

Composition:

1. D-Glucose 
2. Polyvinyl acetate (aqueous emulsion)
3. Fumed silica
4. Sodium Benzoate 

5. Disodium EDTA
6. Food Pigment Red No.6
7. Antifoaming agent
   (Polyethylene Glycol 4000)

How to Perform a Control Test

NOTE :

   

1. Be sure that you are in Control Solution Mode so that the test result will not be stored 

in themeter memory.

           2. DO NOT APPLY THE CONTROL SOLUTION DIRECTLY TO THE TEST STRIP! 

Overdosed solution may give inaccurate result.

              3. Repeat test if test result falls outside the control range stated on the test strip label. If 

subsequent test remains to produce unacceptable result, the meter or test strip may 

be faulty. DO NOT use the system. Contact us or your local distributor for help. 

1. 

Insert Test Strip :

 Insert a new test strip into the strip slot, 

the meter will activate.

2. 

Mark as a Control Solution Test : 

After the blood 

symbol(  ) appears, press M BUTTON and a CONTROL 

SOLUTION TEST SYMBOL(  ) shows up, now you are in 

the Control Solution Mode. The meter will not store your 

test result in the memory when you preset the test as a 

control solution test. If you decide not to perform a control 

solution test, press M BUTTON again and the CONTROL 

SOLUTION TEST SYMBOL (  ) will disappear.

3. 

Squeeze a drop of Control Solution : 

Shake control 

solution bottle well. Remove the cap. Squeeze bottle, 

discard the first drop and wipe off the dispenser tip with 

a clean tissue paper or cotton swab. Squeeze a drop on 

clean bottle cap/fingertip /non-absorbent surface.

4. 

Apply Control Solution : 

Apply the drop to the opening 

of the strip absorbent channel until the confirmation 

window is filled. The meter begins to count down.

 

5. 

Check if the test result is in range: 

After the meter 

counts down, the test result shows up. Compare the test 

result with the range printed on the test strip vial. The 

result should fall within the printed range. 

6

.

The following is a summary of some display messages and symbols. These messages 

help to identify certain problems but do not appear in all cases when a problem has 

occurred. 
Improper use may cause an inaccurate result without producing an error message.
In the event of a problem, refer to information under ‘‘action to take’’

DISPLAY

DESCRIPTION

ACTION TO TAKE

Display check

If some parts of the display are not 

working. Contact your local distributor 

for help.

Moving Strip

The meter is waiting for test strip to be 

inserted.

Blinking Blood

The meter is ready for blood applying 

into test strip.

Deleting memory

Deleting is complete.

Test result is higher than
600 mg/dL (33.3 mmol/L).

If this is not confirmed by the way you 

feel, review proper testing procedure 

and perform a control test. Repeat blood 

test, if the display still appears, please 

call medical assistance immediately.

 

Test result is lower than
20 mg/dL  (1.1 mmol/L).

If this is not confirmed by the way you 

feel, review proper testing procedure 

and perform a control test. Repeat blood 

test, if the display still appears, please 

call medical assistance immediately.

B

     System Specifications

Model Name

BG-101

Assay Method

Electrochemical biosensor

Test Sample

Capillary Whole Blood

Test Result

Referenced to plasma glucose value

Alternative Site Testing

YES (palm, forearm)

Sample Size

0.3 μL

Measuring Time

6 seconds

Measuring Range

20 – 600 mg/dL (1.1 – 33.3 mmol/L)

Acceptable Hematocrit Range 20~60%

Operating Condition

10°C~40°C (50°F~104°F), between 10-85% R. H.

Storage/Transportation Condition 4°C~40°C (39°F~104°F), between 10-85% R. H.

Memory Capacity

120 test results with time and date

Average Calculation

7, 14, and 28 days

Power Supply

One 3-volt Lithium Battery (battery type CR2032)

Battery Life

Approximately 1,000 tests

Automatic shut-off

In 2 minutes

Dimensions

80 mm(L)x 48 mm(W) x 15 mm(H) 

Weight

37 g

      Memory Recall

The Medicare System-LiS automatically stores 120 test results, letting you review 

them in order from the most recent to the oldest. The meter also calculates and 

displays 7, 14 and 28-day averages. You can review the individual or average test 

result by entering the memory mode.

STEP 1. Enter the Memory Mode

While the meter is turned off, press M BUTTON to turn on the 

meter. When the strip symbol blinks in the display, press M 

BUTTON again to enter memory mode.

STEP 2. Recalling Average Test Results

When entering the memory mode, the 7-day average will 

appear. If you continue to press the M BUTTON, the 14-day 

and 28-day averages will appear in order. 

STEP 3. Recalling Individual Test Results

After 28-day average, the most recent test result with date 

and time will be shown. Press M BUTTON once and the next 

most recent test result will appear. Each time you press and 

release the M BUTTON, the meter will recall up to your last 

120 test results in order. When the memory is full, the oldest 

result is dropped as the newest is added. After reaching the 

last set of result, the meter will turn off automatically.

STEP 4. Exit the Memory Mode

If you want to exit the memory mode before reaching the last 

result, press and hold M BUTTON for three (3) seconds to 

turn off the meter.

NOTE :

 

1. The control solution results are NOT stored in the memory if you preset the 

control test correctly. (Please refer to the “CONTROL SOLUTION TEST” Section 

for details). The list of past results and the result average are for blood glucose 

results only.

2. 

When using the meter for the first time, ‘‘----‘’ is displayed. When you recall the 

test results or review the average result, it means that there is no test result in 

memory.

3. The averages are calculated from your latest result obtained during the last 7, 14 

and 28 days. 

4. Anytime in memory mode, you can press M BUTTON for three (3) seconds to 

exit and turn off the meter.

5. While in the memory mode, if you leave the meter alone without any action for 

one (1) minute, the meter will turn off automatically.

5

.

     Understanding Your Test Result

The normal blood glucose range is below 100 mg/dL (5.6 mmol/L) for a 

fasting, non-diabetic adult, but less than 140 mg/dL (7.8 mmol/L) two hours 

after meals.* Consult your healthcare professional to find out your target blood 

glucose value.

If your blood glucose result seems unusually high or low, or inconsistent with 

your previous results, check the following:  

1.  Was the blood sample applied immediately to the test strip after removing it 

from the vial?

2.  Was the volume of the blood sample sufficient? 

3.  Was the test strip vial cap tightly sealed?

4.  Was the test strip used before the expiration date? 

5.  Were the test strips stored away from extreme temperatures in very cold or 

hot weather or from areas of high humidity? 

Then run a control test with Medicare control solution. If the control test result 

is within the acceptable range, review testing procedure and repeat your blood 

glucose test with a new test strip. If your blood glucose value is still inconsistent 

with your previous results, glucose trend, or how you feel, contact your doctor 

immediately for help.

REFERENCE:
*American Diabetes Association (http://www.diabetes.org/diabetes-basics/prevention/pre-diabetes/diagnosis.html)

4

.

DISPLAY

DESCRIPTION

ACTION TO TAKE

 

Battery is weak.

Replace battery soon.

 

Battery is dead.

Replace battery now.

Maybe:

1. Used strip or 

moistened strip is 

inserted. 

2. Meter is defective.

You have to:

1. Repeat test with a new test strip.

2. Contact your local distributor for 

help.

Temperature is out of the 

operating range.

The meter is not working. Move to an 

area with temperature between 10ºC to 

40ºC (50ºF - 104ºF) and wait at least 30 

minutes. Do not artificially heat or cool 

the meter.

No responses 

when strip is 

inserted into 

the meter

Maybe:

1. Battery is dead.

2. Wrong strip be 

inserted.

3. Meter is defective.

You have to:

1. Replace battery.

2. Insert the test strip correctly.

3. Contact your local distributor for 

help.

No responses 

when blood 

sample is 

applied to the 

strip 

Maybe:

1. Blood sample is not 

sufficient.

2. Meter is defective.

You have to:

1. 

Repeat test with sufficient 

sample. 

2. Contact your local distributor for 

help.

     Performing Blood Test

1. 

Wash Your Hands and the Puncture Site :

 Wash 

your hands in warm, soapy water. Rinse and dry 

completely. Warm your fingers to increase blood flow.

2.  

Insert Test Strip : 

Remove a new test strip from vial. 

Be sure to tightly replace vial cap after removing test 

strips. Insert test strip immediately into strip slot as 

illustrated. The meter turns on automatically. When the 

blood symbol blinking, you are ready to perform a test. 

3. 

Select and Lance a Puncture Site

*For Fingertip

Hold the prepared lancing device firmly against the 

side of your fingertip. Press release button. 

*For AST

Please refer to the “About Alternative Site Testing 

(AST)” Section. Please consult your healthcare 

professional before obtaining blood from site other 

than your fingertip.

4. 

Obtain a Blood Sample : 

Gently massage your finger 

or puncture site to obtain the required blood volume. 

To perform the test, you need only 0.3 μL of blood 

sample. Do not smear the blood sample. To obtain 

best accurate result, wipe off the first drop of blood and 

gently squeeze another drop of blood.

5. 

Apply Blood Sample : 

Apply the blood sample to the 

opening of absorbent channel of test strip until the 

confirmation window is fully covered with blood. Then 

the meter begins to count down.

 

6. 

Read Your Result : 

   The meter will display your blood glucose value after 6 

seconds.

7. Remove Strip to Turn Meter Off : 

   Your blood glucose result is automatically stored in the 

meter memory. Turn the meter off by removing the test 

strip.

8. 

Secure the Used Lancet : 

Twist off the lancing device 

cap, and push the exposed tip of the lancet into its 

protective cap.

9. 

Discard the Used Lancet : 

Slide the lancet ejector 

forward and dispose the lancet. Discard the lancet 

according to your safety regulations. 

     * Do not reuse lancets.

Incorrect

Correct

3

.

90°

To avoid the meter and test strips getting dirt, dust or other contaminants, please 

wash and dry your hands thoroughly before use.

Cleaning

Your meter does not require special maintenance. As long as no blood or control 

solution comes in direct contact with the meter, there is no special cleaning 

required.
To clean the meter exterior, wipe with a cloth moistened with tap water or a mild 

cleaning agent, then dry the device with a soft and dry cloth. Do not flush with 

water.
Do not use organic solvents to clean the meter. Your meter is a precision 

instrument. Please handle it with care.

Storage

1. Meter Storage

* Storage condition: -20ºC~50ºC (-4ºF~122ºF), below 90% relative humidity.

* Avoid dropping and strong impact.

* Avoid direct sunlight and humidity.

2. Strip Storage

* Storage condition: 4°C~40°C (39°F~104°F), and 10~85% relative humidity. Do 

not freeze.

* Store your test strips in their original vial only. Do not transfer to other 

container.

* Store test strip packages in a cool and dry place. Keep away from direct 

sunlight and heat.

* After removing a test strip from the vial, immediately replace the vial cap and 

close it tightly.

* You may touch the test strip anywhere with clean, dry hands when removing it 

from the vial or inserting it into the meter.

* Use each test strip immediately after removing it from the vial.

* Do not bend, cut, or alter a test strip in any way.

* Keep the strip vial away from children since the cap and the test strip can be 

a potential choking hazard. If swallowed, please seek medical assistance 

immediately.

3. Control solution storage

* Store the control solution in a cool, dry place between 4°C (39°F) and 30°C 

(86°F).

  Caring For Your Meter And

                            Test Strip

8

.

7

.

4

5

Reviews: