15.Troubleshooting
Description
of faults
Solutions
Inaccurate
measurement
1.Check the measurement object. The measurement performance
cannot be guaranteed in the case of forehead blowing, water
dripping, sweating, applying cosmetics, and applying antipyretic
paste;
2. Check the operating environment. Environmental changes have a
greater impact on the measurement. If the environmental
temperature changes too much, or the instrument has just measured
the temperature of an ultra-high object and directly converted to the
measurement of a low-temperature object, the measurement
difference will occur. Leave it for about 10 minutes to obtain a new
test heat balance before using it;
3.The best operating environment of the product is 16 ºC -35ºC ,
check whether the current environmental temperature is too low or
too high.
4.Check the distance of measurement (3-5cm).
Screen
displays
"LO" or
"HI"
1.The temperature of the tested object is lower than 34.1ºC, the
screen displays LO;
2. The temperature of the tested object is higher than 42.9ºC, the
screen displays HI;
3.Check whether the internal parameter settings are disordered.
No response
to button
Reload the battery.
No display
or display
abnormally
Reload the battery.
Shut down
immediately
after power
on
Check whether the battery power is sufficient or replace the battery.