EasyOne Pro™ / EasyOne Pro™ LAB
117
Specifications
10 Specifications
Dimensions
270 x 335 x 270 mm
3
(h x w x d)
Weight [kg]
< 8 kg, LAB < 9 kg
Flow/volume
measurement
Accuracy, volume:
±2% or 0.050 l
Accuracy, flow:
±2% or 0.020 l/s
(except PEF)
Accuracy, PEF:
±5% or 0.200 l
Accuracy, MVV:
±5% or 5 l/min
Resolution, volume
>1 ml
Resolution, flow
>4 ml/s
Measuring range, volume ±12 l
Measuring range, flow
±16 l/s
Resistance
<1.5 cm H
2
O/l/s at 12 l/s
Molar mass
analyzer
Type
ultrasonic transit time
Measuring range
9 to 50 g/mol
Accuracy
<0.02 g/mol
Resolution
0.01 g/mol
Carbon
monoxide
analyzer
Type NDIR
Measuring range
0 to 0.35%
Accuracy
±0.001%
Resolution
0.0001%
CO2 analyzer
(LAB only)
Type nondispersive
infrared
Measuring range
0 to 15%
Resolution
0.005%
Accuracy
from 0 to 5%: ±0.05%
from 5 to 10%: ±0.20%
>10%:
not specified
Mouth
pressure
transducer
Measuring range
±100 mbar
Accuracy
±1%
Predicted
values
(spirometry)
NHANES III, Knudson_83, Knudson_76, Crapo, Mor-
ris, Hsu, Dockery (Harvard), Polgar, Gutierrez
(Canada), Eigen, Pereira, Pereira 2006/2008 Chile
2010/1997, ERS (ECCS / EGKS), Zapletal, Stanojevic
2009 (GLI), Rosenthal, Austria 1988, Austria 1994,
Sapaldia (Swiss), Roca (Spain, SEPAR), Vilozni 2005,
Hedenström (Sweden), Gulsvik (Norway), Berglund,
Birath (Sweden), Langhammer (Norway), Finnish,
Nystad, Hibbert, Gore, Crockett, Ethiopia, Asia 1-4,
JRS 2001, Fukuda Standard (see