Medical Freezer MPF1.8L 80 Operation Manual Download Page 9

- 7 -

Ultra-low temperature freezer: fasten the door lock and move it lightly.

Remove all packaging components (including protective foam inside the cabinet).

Please check the attachments in the device according to packing list.

The product should be cleaned before use.

Environmental requirements:

Indoor use

● The installing surface must be fixed, level, non- flammable and capable of withstanding the weight of the
freezer when it is in operation.
● It should be placed away from direct sunlight and away from heat sources, and the ambient temperature
should not be higher than 32

.

● Leave more than 30cm space around for easy ventilation and heat dissipation.
● Do not place freezer in a too cold environment.
● Do not place freezer in heavy moisture or easy-splashing water places.
● No strong electromagnetic interference around which will affect the normal operation of the control system,
and even will directly damage the system.

Normal working conditions:

a. Ambient temperature: 10 ° C ~ 32 ° C
b. Relative humidity: ≤ 80%
c. No strong vibration and no corrosive gas around.
d. No direct sunlight or any other cooling or heating sources.
e. Keep away from factors that may cause damage or interference to the device.

Safety caution

● Power supply voltage (power adapter): The device is equipped with a wide- voltage power adapter as
standard. It uses 220V/50HZ AC power supply or 110V/50Hz AC power supply. When the voltage
compensation function is not selected, if the voltage is lower than 100V or higher than 240V you need to install
a suitable automatic voltage regulator for use;
● Power supply voltage (host): 24v DC power.
● Do not lengthen the length of the power cord. If it is really necessary to lengthen, be sure to use a copper
conductor of 2.5mm2 or thicker. The cross- sectional area of the copper conductor in the wall connected to the
power socket must be above 4mm2.

The freezer working system: the chiller freezer will never stop under the normal running state.

VI. Usage Instructions

1. Display Panel Function Descriptions

Summary of Contents for MPF1.8L 80

Page 1: ...OPERATION MANUAL Applicable product model MPF1 8L 80 Medical Freezer Ltd Info medical freezer...

Page 2: ...s 7 1 Display Panel Function Descriptions 7 2 First Initial Starting 8 3 Function Setting 8 VII Defrosting Unused and Maintenance 10 1 Defrost 10 2 Stop Using 10 3 Maintenance 10 VIII After sales Serv...

Page 3: ...Freezer Only NANBEI or agents authorized by NANBEI could maintain this device If the operator encounters a situation not mentioned in this manual please contact the authorized agent of NANBBEI Biomed...

Page 4: ...2...

Page 5: ...3...

Page 6: ...4...

Page 7: ...s Be extremely careful not to use volatile solvents such as gasoline to wipe plastic rubber parts After running for a period of time in the ultra low temperature freezer a layer of frost will form on...

Page 8: ...the product After removing the cold storage block the product can still work normally but the product performance will be reduced Note The inner cover of the ultra low temperature freezer can be effec...

Page 9: ...damage the system Normal working conditions a Ambient temperature 10 C 32 C b Relative humidity 80 c No strong vibration and no corrosive gas around d No direct sunlight or any other cooling or heati...

Page 10: ...will appear on the screen The initial password is 000000 iii Click the switch button to operate the device 3 Function Setting Regular display In normal use the current temperature is displayed as the...

Page 11: ...word is entered the left curve number will be displayed The alarm code information is as follows Do not put the items in the ultra low temperature freezer that has just been connected to the power sup...

Page 12: ...oth to dry the water around the cabinet especially the part where the door seal is in direct contact with the cabinet to avoid freezing the door seal at low temperature 3 Maintenance Clean and maintai...

Page 13: ...e otherwise the hot air enters the inside of device and the negative pressure caused by the sharp drop in temperature causes the door to be difficult to open Once there is abnormal alarm when turn on...

Page 14: ...ators 1 Main Technical Parameters Model MPF 1 8L 80 Temp range 40 C to 80 C Volume 1 8 L Style vertical with handle External dimensions width depth height 245 282 496mm Internal dimensions width depth...

Page 15: ...rrent A 0 46A Power 80W 2 Packing List Name Operation manuel Quantity 1 3 Electrical Schematic Diagram Due to the rapid improvement of the product please refer to the technical data and electrical sch...

Reviews: