Medical Freezer MPF 20L22 Operating Manual Download Page 3

Operating Manual - MPF

www.medical-freezer.com

[email protected]

3

1.  Manual Notifications

WARNING !

Safety instructions:

 Improper operation may cause fatal or serious injury, and damage to the 

refrigerator.Safety instructions related to current or voltage. Failure to follow this instruction may 

lead to injury or damage to the refrigerator and impair the function of the product.

CAUTION !

Safety instructions:

 failure to observe this instruction can lead to injury.

2.  Safety Instructions

WARNING !

 

The manufacturer hereby declares not responsible for damage caused by the 

following actions.

•  Overvoltage and mechanical breakage lead to refrigerator damage. 

•  Retrofitting the refrigerator without the permission of the manufacturer. 

•  Failure to follow the manual instructions.

WARNING !

 

Warning - Fatal Hazard from Electric Shock. 

•  Never touch the damaged power cable directly with your hands, especially when it is connected 

to AC mains supply. It’s very dangerous!

•  Disconnect the refrigerator and other power consuming devices from the battery before connecting 

the quick charging device.

•  Check that the voltage specification on the type plate corresponds to that of the energy supply.

•  Do not try to open the circuit part of the product.

CAUTION !

•  If you are operating your refrigerator on board a boat from a mains connection of 100-240V, you 

must install a residual current circuit breaker between the 100-240V mains and the refrigerator.

•  Seek advice from a trained technician. Battery contains acid and corrosive substances. Careful 

with using.

•  Do not store any corrosive substances. Do not place obstructions in or near the compressor 

refrigeration system of the product. Do not use electrical devices inside the refrigerator.

•  If this device’s power cable is damaged, it must be replaced by the manufacturer or a similarly 

qualified person in order to prevent safety hazards.

Summary of Contents for MPF 20L22

Page 1: ...zer Operating Manual MPF 20L22 30L22 50L22 Medical Freezer LTD Paul Street 86 90 EC2A 4NE London Greater London England United Kingdom Email info medical freezer com Tel 44 1865 60 0437 www medical fr...

Page 2: ...ore Initial Use 4 3 2 Power on Operation 5 3 3 Battery Protector 6 4 Use of Refrigerators 6 4 1 Power On Off 6 4 2 Setting Temperature 7 4 3 Running Mode Setting 7 4 4 Bluetooth Connection Setting 7 4...

Page 3: ...touch the damaged power cable directly with your hands especially when it is connected to AC mains supply It s very dangerous Disconnect the refrigerator and other power consuming devices from the bat...

Page 4: ...n the refrigerator is working but no influence to the refrigeration function The raw materials used in the production process all meet the restrictions and requirements of all kinds of laws and regula...

Page 5: ...voltage loss from affecting the cooling effect of the product the power cable should be as short as possible Always ensure that the power cable is straight through Therefore avoid the use of addition...

Page 6: ...effectively protect the battery The relevant start voltage and disconnect voltage data as below Hazard warnings In order to protect your battery from being completely depleted and affecting other use...

Page 7: ...The C F max eco H M L of the LCD display start to flash and you can set the below functions Press the setting button the battery protection level H M L displayed alternately you can set the battery v...

Page 8: ...in the cooling device to keep cool Defrost the refrigerator regularly Or defrost the refrigerator once a layer of ice forms Avoid unnecessarily low temperatures Do not leave the lid of the refrigerato...

Page 9: ...ard cleaning tools such as brushes for cleaning Please wipe the refrigerator with a dry cloth after cleaning 5 2 Maintenance WARNING The effect of damaging the refrigerator Humidity can form frost in...

Page 10: ...ault indicator flashes it means the DC power supply voltage is lower than the currently set Battery Voltage Protection gear The battery needs to be charged or set the Battery Voltage Protection gear l...

Page 11: ...edical freezer com 11 Note In case of the above 2 3 4 and 5 failures you can switch off the device and restart it after sufficient time If the fault persists you need to contact an authorised repair c...

Page 12: ...Operating Manual MPF www medical freezer com maintenance medical freezer com 12 7 3 Abnormal Noise...

Page 13: ...hnical modifications and delivery options This refrigerator is a non fluorine refrigerator and the refrigerant is R134a Restriction of Hazardous Substances Directive RoHS To meet the laws regulations...

Page 14: ...Operating Manual MPF www medical freezer com maintenance medical freezer com 14...

Reviews: