QUESTION:
I’ve followed all of the installation instructions and my pendant still doesn’t
work. What can I do?
ANSWER:
1. Check the base station connections. Ensure the AC/DC adapter is plugged into
a live power outlet and also into the base station. The GREEN POWER indicator
on the top of the base station should be illuminated. If power outlet is controlled
by a wall switch, make sure that the switch is in the ON position.
See STEPS 2-4
of
INSTALLING YOUR
MediPendant™
on pages 6 and 7.
2. Confirm that the telephone line coming from the wall jack is plugged into the base
station in the YELLOW jack marked PHONE LINE IN FROM WALL OUTLET.
See STEP 5 of
INSTALLING YOUR
MediPendant™
on page 8.
3. Press and hold the BATTERY/SYSTEM TEST BUTTON on the bottom of
the pendant for 3 seconds and release.
See Figure 17A on page 17.
The unit
should announce battery status followed by, "System check, please stand by.”
The pendant will turn on and send a test signal to the base station to confirm
all functions are working properly.
If all functions are normal, the pendant with
announce, ”All functions are OK.” If not, the pendant states,”System test fails.”
If test fails and you have tried answers 1 and 2 above, please contact your local
dealer at the number below.
If nothing happens, the most likely cause is that
the battery is dead or has been installed improperly.
See
INSTALLING/REPLACING THE BATTERY
on page 18.
PENDANT/BASE STATION SYNCHRONIZATION
The pendant and the base station are connected by a wireless radio signal. They
need to be synchronized to work. All systems are synchronized prior to
shipment.
However, in some cases, you may need to re-synchronize your
system. Press the button on the base station marked SILENT/SET UP for three
seconds. The RED PHONE LINE INDICATOR LED will illuminate and flash. Press
and hold the LARGE GREY BUTTON on the pendant simultaneously with the
BATTERY/SYSTEM TEST BUTTON on the bottom of the pendant for a few
moments until the green LED on the right side of the pendant blinks. Once
green flashing begins, release buttons.
Once the pendant and the base unit are
synchronized, the green LED on the pendant and the red LED on the base
station will go out and your pendant will now be synchronized to the base
station. If done properly, you will hear the voice announcement, “System check.
All functions
are okay.”
19
OWNER’S MANUAL
Model #MED01
TROUBLESHOOTING
IF YOU STILL HAVING DIFFICULTY WITH YOUR
MediPendant™
SYSTEM,
PLEASE CALL THE CUSTOMER SERVICE DEPARTMENT AT
Phone #: _________________________________________________