Guidance and manufacturer *s declaration-electromagnetic immunity
The infrared thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the infrared
thermometer should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment
-guidance
Conducted RF
IEC 61000-
4-6
3Vrms
150kHz to
80MHz
N/A
Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
infrared thermometer
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
seperation distance
d=1.2 P
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-
4-8
3A/m
3A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment
NOTE UT is the a.c. mains voltage prior to application of the test level.