Medexpert CBPM24 User Manual Download Page 13

 

 

13

 

 

calibration

 

takes

 

place,

 

setting

    

the

 

zero

 

pressure

 

(see

 

adjacent

   

figure)

 

 

CAL 0 

After

 

that,

 

the

 

measurement

 

starts

 

by

 

the

 

inflation

 

of

 

the

 

cuff,

 

signaled

 

on

 

the

 

display

 

(see

 

adjacent

 

figure).

 

          87 

 
 

The

 

device

 

checks

 

the

 

placement

 

of

 

the

 

cuff

 

during

 

inflating.

 

If

 

the

 

cuff

 

on

 

the

 

arm

 

is

 

too

 

loose

 

or

 

not

 

the

  

proper

 

size

 

of

 

cuff

 

has

 

been

 

chosen

 

(e.g.

 

it

 

is

 

bigger),

 

the

 

following

 

sign

 

will

 

be

 

on

 

the

 

display

 

accompanied

 

by

 

a

 

beeping

 

signal.

 

Check

 

the

 

cuff

 

and

 

its

 

tightness

 

and

 

repeat

 

the

 

blood

 

pressure

 

measurement.

 

 

CUFF 

The

 

deflation

 

of

 

the

 

cuff

 

is

 

shown

 

by

 

the

 

adjacent

 

figure

  

 

 

       69 

After

 

this

 

process

 

the

 

device

 

shows

 

the

 

systolic

 

and

 

diastolic

 

BP

 

values

  

Then

 

the

 

pulse

 

rate

 

is

 

shown

 

on

 

the

 

display,

 

and

 

the

 

device

 

stores

 

all

 

the

 

measured

 

data,

 

including

 

the

 

date

 

and

 

exact

 

time.

 

 

128/96 

PUL 68 

At

 

any

 

time

 

during

 

a

 

reading

 

the

 

patient

 

can

 

terminate

 

a

 

reading

 

by

 

pressing

 

once

 

the

 

single

 

button.

 

A

 

termination

 

symbol

 

will

 

appear

 

on

 

the

 

display

 

for

 

10

 

sec

 

(see

 

adjacent

 

figure).

 

Then

 

the

 

time

 

will

 

appear

 

and

 

the

 

units

 

ready

 

for

 

measurement,

 

for

 

manual

 

and

 

programmed

 

mode.

 

 

OFF 

 

 

2.4.1.2.

 

Two

 

short

 

pushes

 

on

 

the

 

button

 

(Pill):

 

allows

 

the

 

patient

 

to

 

keep

 

his

 

"electronic

 

diary"

 

concerning

 

taking

 

his

 

(antihypertensive)

 

medication.

 

After

 

taking

 

his

 

medicine,

 

two

 

short

 

pushes

 

on

 

the

 

button

 

stores

 

the

 

date

 

and

 

time

 

in

 

the

 

memory.

 

During

 

a

 

day,

 

it

 

is

 

possible

 

to

 

store

 

additional

 

pill

 

consumptions.

 

By

 

downloading

 

all

 

the

 

data

 

from

 

the

 

device

 

to

 

the

 

physician's

 

PC,

 

he

 

will

 

be

 

able

 

to

 

monitor

 

the

 

medication

 

intake

 

and

 

therefore

 

the

 

compliance

 

of

 

the

 

patient.

 

 

Pill 

 

If

 

the

 

memory

 

of

 

the

 

device

 

is

 

full,

 

this

 

will

 

appear

 

on

 

the

 

display.

 

 

FULL 

Summary of Contents for CBPM24

Page 1: ......

Page 2: ...2...

Page 3: ...Installing the Device 11 2 4 Operations Instructions 12 2 4 1 Functions of the button on the device 12 2 4 1 1 One short push of the button 12 2 4 1 2 Two short pushes on the button 13 2 4 1 3 Three...

Page 4: ...ate measurement frequencies can be set for the active daytime passive night time and for a third special period Four different types of blood pressure measuring plans with different measuring frequenc...

Page 5: ...ir respective companies 1 1 Contents of the manual This manual helps you in setting up and starting to use the Medexpert CBPM24 device Attention Before first use please read and understand this docume...

Page 6: ...5 7 Budapest H 1037 Hungary Phone 36 1 250 4490 Fax 36 1 433 3795 Web www medexpert hu E mail arterialcare medexpert hu 1 4 General Information warnings and precautions We strongly suggest that you c...

Page 7: ...fied service Only use with cuffs supplied by Medexpert Ltd Use of cuffs supplied by a third party can lead to erroneous measurement results Confirm blood pressure measurement with auscultation when er...

Page 8: ...2 The Medexpert CBPM24 device 2 1 Explanation of symbols 1 Name of the device 2 Function button Offering five menu options see section 2 4 1 3 Command symbols see section 2 4 1 4 LCD display 5 Air con...

Page 9: ...1 The classification of the protection against electric shock Classification patient s side CF 12 Calling the attention to read thoroughly the present User s Manual 13 Certification mark guaranteeing...

Page 10: ...batteries 3 different cuffs Medexpert TensioWin software on CD User manual IrDA adapter For the 3 different size cuffs the dimensions are Bladder Dimensions Sleeve Dimensions Arm Circumference Cuff 01...

Page 11: ...ore the clock time is held and resetting the time is unnecessary between battery changes If you do not intend to use the device for a long period of time remove the batteries and store them in a cool...

Page 12: ...al will be between 5 4V 4 4V The nominal potential is 4 8V If the voltage is under 4 4 V the batteries need replacing The replace battery message will be displayed If long life batteries are used to p...

Page 13: ...e stores all the measured data including the date and exact time 128 96 PUL 68 At any time during a reading the patient can terminate a reading by pressing once the single button A termination symbol...

Page 14: ...ill be continued OFF 2 4 2 Data transmission The device downloads all the stored data to the physician s PC via infrared communication The information loaded consists of the systolic and diastolic blo...

Page 15: ...as interrupted due to the weakness of the battery 31 The cuff is not connected to the device 32 The cuff tube is broken or something got into the tube e g water 33 The cuff or device is leaking 34 The...

Page 16: ...ff the tube of the cuff should be inserted into the black opening on the left side of the device Attention Please take care of the connection of the cuff because it should not be too loose it should n...

Page 17: ...Hg Static accuracy 3 mmHg or 2 of the measured value Measuring accuracy Systolic 94 out of 99 comparisons were within 5 mmHg 95 in case of 33 out of 33 patients 2 comparisons out of 3 were within 5 mm...

Page 18: ...18...

Page 19: ...19...

Page 20: ...20...