Guidance and manufacturer's declaration –
electromagnetic immunity –for all EQUIPMENT and
SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)IEC
61000-4-2
±8 kV
contact
±15 kV
air
±8 kV
contact
±15 kV
air
Power
frequency
(50/60Hz)
magnetic
field IEC
61000-4-8
30 A/m
30 A/m
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.