Guidance and manufacturer’s declaration – electromagnetic immunity
This breast pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the breast pump should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the breast pump, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommend separation distance
d = 1,2√P
d = 1,2√P 80 MHz - 800 MHz
d = 2,3√P 800 MHz - 2.7 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol.
10 V/m
(minimum)
10 V/m (minimum)
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Radiated RF
IEC 61000-4-3
6V
3 V
0,15 MHz – 80 MHz
6 V in ISM and
amateur radio bands
between 0,15 MHz
and 80 MHz
80% AM at 1 kHz
Conducted RF
IEC 61000-4-6
12. EMC technical description (cont.)