135
Electromagnetic immunity
The electric breastpump Mini Electric is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Mini Electric should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
– guidance
Electrostatic
discharge ( ESD )
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line( s ) to
line(s)
± 2 kV line( s ) to
earth
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)