8
2.1.5 Assembly
Assembly is performed in the reverse sequence to dismantling. The
following points should also be noted:
– Sequence of cable routing from top to bottom: mains lead from
SPS, potentiometer (RV1), switch, LED (D1)
– The SPS must be placed right at the rear of the case and the
cables must be pressed fully into their holders
– Insert potentiometer (RV1) in spacing piece (25) in positioning pin
from inside to outside The potentiometer must lie flat on the
spacing piece Place nuts on from outside and tighten
– Turn potentiometer clockwise to stop, fit knob (26 + 27) and turn
clockwise until line is at 7 Tighten screw
– Before fitting the back check whether the internal parts of the
pumping arm (2) do not touch anywhere
– Connections on right-hand side of mains terminal strip (KL1) from
top to bottom: Power cord no 1, power cord no 2 and SPS
brown, SPS blue
– Connections on left-hand side of mains terminal strip from top to
bottom: black, blank, white
– Fitting of the cover insert (7) is simplified if the pumping arm (2) is
not swung in completely, but only set to the final position together
with the cover insert (7)
Breastpump Systems
Breastmilk Initiation System for Symphony or Lactina
67340S
Symphony / Lactina
67116
67116-06
67116S
67340S
Lactina
67094
67094-06
67094S
Symphony / Lactina Rental Kit
67206-03
Summary of Contents for lactrina
Page 2: ...service instructions lactina electric breastpump ...
Page 7: ...4 2 1 2 Spare Parts List ...
Page 9: ...2 1 3 Electric Circuit Diagram 6 IO I J1 ...
Page 19: ......
Page 20: ......