SET UP AND OPERATION:
1. Carefully remove the Aspirator IV from the shipping carton. Be sure
to save the box and packing material for convenient return to Medco.
2. Read all instructions carefully before operating the unit.
3. After you have removed the Aspirator IV, open the accessory box and
locate the power cord and foot switch. Attach the foot switch to the
Aspirator IV. Plug the power cord into the Aspirator IV and then into
the power outlet.
4. Locate the vacuum gauge and screw into the gauge fitting on top of
the Aspirator IV.
5. Turn the Power switch and the Foot Switch ON.
6. To test the unit, place a finger over the main suction hose coming
directly out of the top of the Aspirator IV. Note the reading of the
vacuum gauge. Keep in mind that vacuum changes according to
elevation. After checking the vacuum, turn the Aspirator IV off.
7. Connect the HEPA Filter to the Aspirator IV using the
6”
silicon tube.
Connect the other end of the HEPA Filter to the permanent canister.
Note: Change the HEPA Filter after 10 procedures or every
24 hours use or replace Filter IMMEDIATELY if ANY
Bodily Fluids/Blood/Fat/Residue is visual within the Filter.