Page 1: ...lasty Since 1979 M A N U F A C T U R I N G ASPIRATOR IV MANUAL PRIMARY MEDICAL DEVICE LABEL Medco Manufacturing 8319 Thora LN Bldg A1 Spring TX 77379 281 379 3100 800 953 2555 281 251 7608 Fax www Med...

Page 2: ...sulation system make the MEDCO ASPIRATOR IV the QUIETEST Aspirator on the market today Product L0020 2000ML Disposable Liner Box 10 Product C0001 2000ML Reusable Hard Canister Each Product B1540 Premi...

Page 3: ...ion system bacterial filter and finally to the pumps that will evacuate the air from the lines creating a vacuum pulling aspirated material in the same direction as the air flow Aspirated material con...

Page 4: ...ister Liner Product L0020 Box 10pcs User s Manual la carte item s NOT INCLUDED AS AN ACCESSORY Listed for disposable s reorder reference Call Customer Service Immediately to report any damaged or miss...

Page 5: ...trol by turning pointer between High and Low pressure is reduced The vacuum gauge in the panel will indicate a pressure change Note The Aspirator IV system has been calibrated at 160 feet above sea le...

Page 6: ...Avoid ingesting aspirated material into the pumps Severe damage can occur to the pumping system upon contact with aspirated material Always disconnect the power supply to the Aspirator IV before serv...

Page 7: ...screw into the gauge fitting on top of the Aspirator IV 5 Turn the Power switch and the Foot Switch ON 6 To test the unit place a finger over the main suction hose coming directly out of the top of th...

Page 8: ...Set Up Operation 1 Place canister on the top panel of the Aspirator IV inside canister black base holder 2 Insert a Quick Fit Liner into the canister and secure the lid 3 Connect the piece of tubing...

Page 9: ...lubricate pumps The oil less components are sealed and packed with optimum level of lubricant requiring no lubrication Do not let liquid or aspirated material enter the pumping system The mufflers sh...

Page 10: ...a Vacuum Pump 1 3 hp piston hi vac 1 ea Muffler MIN Tubing Internal PVC la carte item s NOT INCLUDED AS AN ACCESSORY Listed for disposable s reorder reference Trouble Shooting No Suction or Trips Brea...

Page 11: ...r can cause aspirated material to overflow into the overflow trap 3 Check for a crack in the canister This can cause the liner to collapse resulting in an overflow situation A collapsing liner bag doe...

Page 12: ...year on parts and labor from the date of shipment Unauthorized handling of the Aspirator IV will void warranty of the unit Any aspirated material found to have been ingested into the pumping system wi...

Page 13: ...o based on the dollar amount will be issued to the account 6 OPENED PACKAGES OR BOXES containing disposable are not returnable for credit or exchange Each package or sales unit specifically states Not...