Alarms and Troubleshooting
44
High
“Di-di-di..di-di…Di-di-di..di-di” is given every
6s.
The red indicator
flashes.
Prompt
“Di-di” is given transiently.
/
After an alarm is triggered, you can tap
Silent
to silence the alarm. Two minutes later, the alarm sound is
given again if the alarm level is middle or high and it still exists.
9.2
Common Faults and Troubleshooting
The following table lists the alarm message, alarm level, alarm cause, and troubleshooting.
Alarm Message
Alarm Level
Alarm Cause
Troubleshooting
No AC power
Low
No external AC or DC
power.
Connect to the external
AC or DC power
immediately.
No Battery
Low
No built-in battery or
built-in battery failure
Install the battery.
BAT Low
Low
Low battery power
Connect to the external
AC
or
DC
power
immediately.
BAT Empty
High
Out of battery power.
Connect to the external
AC
or
DC
power
immediately.
Infusion End
High
The VTBI or infusion
time in
Infusion Set
is
reached.
Tap
Stop
to eliminate the
alarm.
Infusion Near End
Low
The infusion will be
completed in the alarm
time.
Wait until the infusion is
completed.
Infusion End KVO
Start
High
The VTBI or infusion
time in
Infusion Set
is
reached and the KVO
process is started.
Tap
Stop
to eliminate the
alarm.
KVO End
High
The KVO process is
started and maintained
for 30 min.
Tap
Stop
to eliminate the
alarm.