Cleaning and disinfection
Product-specific safety notes
GA 5752 2135 GB 18
27
6
CAUTION!
Property damage due to sterilisation!
Do not sterilise the product.
DANGER!
Risk to patient!
Oversuction of products results in them no longer being functional. There is
considerable risk to the patient if the tapping unit is not cleaned properly after being
exposed to oversuction, as safety equipment could be clogged.
After oversuction, products must be dismantled and cleaned thoroughly by
authorised service staff.
CAUTION!
Property damage!
Using non-colour-fast surgical drapes can cause discolouration of surfaces.
Only use colour-fast surgical drapes.
NOTE
For the cleaning and disinfection of version B, disconnect the connection tube with
the NIST screw connection from the tapping unit.