7
2. Safe Place
1) Avoid damaging the plug or cables when moving the patient or bed.
2) Ensure that the cables are safe and tidy and not caught in the bed mechanism or resting on the floor.
3) Keep sharp objects or strong acid and alkali chemical agents away from the mattress.
4) Fuse will blow to protect the device when electric short circuit is caused by the current loading.
3. Cautions - Keep the system away from fire, avoid electric shock or any danger that will cause injury to the user.
1) Only use the system under the supervision of a nurse or caregiver that understands how to operate this device.
2) Do not use any parts that did not come with this unit or supplied by MedaCure or authorized dealer.
3) Ensure that the mattress, cover, cables and power outlet are not in contact with fire, heat or strong acid and alkali chemical
agents.
4) Do not modify the pump and mattress without authorization of the manufacturer and do not insert objects
into the control unit or mattress.
6) Please plug the control unit into the ground power supply.
7) Do not use the product in case of the following:
1 the plug or the cable is damaged.
2 the system sounds abnormal or vibrates intensely.
3 the control unit fell on the floor and caused damaged to the housing
4 the control unit fell into water or other fluid.
HOW TO USE
NOTE
In case the pressure is constantly low, check for any holes on the interchangeable air cells or on the
connection tubes. If necessary, contact your dealer to exchange the air cells or connection tubes/hoses.
HOW TO USE THE CPR DEFLATOR & INFLATOR
In case of emergency and/or if rapid deflation is necessary, rotate the CPR valve to deflate, as noted below:
1) Connect the CPR to the mattress -
Connect the 3 prongs on the CPR Valve
to the CPR hose on the mattress.
Leave the CPR Valve connected to hose in case of future use
.
Position on the
Pressure knob
Weight
Pressure
1
80 LBS
20~24mmHg
2
80~115 LBS
24~28mmHg
3
115~150 LBS
28~32mmHg
4
150~185 LBS
32~36mmHg
5
185~220 LBS
36~40mmHg
6
220~255 LBS
40~44mmHg
7
255~290 LBS
44~48mmHg
8
290~325 LBS
48~52mmHg
9
325~360 LBS
52~56mmHg
10
360~400 LBS
56~60mmHg
CPR Deflator –
How to deflate the mattress using the CPR valve