User Manual
FOR MODEL # CZ36 & CZ36-BIP
8” REPLACEMENT CELL-ON-CELL MATTRESS
© 2016
MADE
IN
CHINA
Page 1: ...User Manual FOR MODEL CZ36 CZ36 BIP 8 REPLACEMENT CELL ON CELL MATTRESS 2016 MADE IN CHINA...
Page 2: ...ORE USE page 6 USE WARNINGS page 6 7 HOW TO USE page 7 HOW TO USE THE CPR DEFLATOR INFLATOR page 7 8 STORAGE page 9 MAINTENANCE page 9 CLEANING AND DISINFECTION page 9 CONDITIONS OF DISPOSAL page 9 Ge...
Page 3: ...ct for any purpose not indicated in this manual Only qualified personnel trained in the treatment and prevention of bedsores should operate this device MedaCure Inc declines all liability for any cons...
Page 4: ...nt otherwise injury or damage may occur 2 Consult with physician or therapist to determine the correct adjustment and the correct use of this device 3 Keep the packed kit away from heat sources 4 SERV...
Page 5: ...ernate keeps mattress firm and always inflated d SEAT FUNCTION To increase pressure in each cell when mattress is in the sitting position COMFORT LOW PRESSURE To increase or decrease airflow for a sof...
Page 6: ...r 5 Fuse Back 8 Filter 9 Main filter 10 Protective cap of filter 11 Hook 1 Pillow 2 Cover 3 Fixing belt 4 Low air loss holes 5 Male quick connector 6 Protective cap of connector 7 Air tubing 8 Patente...
Page 7: ...e of a sudden power outage The mattress will temporarily retain air even when the pump is turned off or during a power outage Please note this only temporary The cover sheet protects the cells against...
Page 8: ...plug the control unit into the ground power supply 7 Do not use the product in case of the following 1 the plug or the cable is damaged 2 the system sounds abnormal or vibrates intensely 3 the control...
Page 9: ...ng 4 Align the air holes as shown to deflate 1 This mattress includes a Quick Inflate Pump and nozzle attachment 2 Attach the nozzle to the pump as shown 3 Align the air holes on the CPR Insert the Qu...
Page 10: ...t by bringing it to a specific recycling point for electric and electronic equipment or others that provide this service SPARE PARTS ACCESSORIES For spare parts and accessories please contact MedaCure...
Page 11: ...emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...
Page 12: ...to line s 2kV line s to earth 1kV differential mode 2kV differential d Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage...
Page 13: ...ngths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment...
Page 14: ...ower of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73...
Page 15: ...peration 10 40 Storage 15 50 Shipping 15 70 Environment humidity Operation 10 90 non condensing Storage 10 90 non condensing Shipping 10 90 non condensing Atmospheric Pressure Operation 700 1013 25hPa...
Page 16: ...alling Patient must be under the constant watch of a caregiver to monitor the patient s movements When the patient needs to transfer out of bed each side can be deflated individually to allow the pati...