Med-link Electronics AM-806VB Instructions For Use Manual Download Page 1

Address: 4th and 5th Floor, Building Two, Hualian Industrial 

Zone, Xinshi Community, Dalang Street, Longhua District, 

518109 Shenzhen, CHINA

Fax: +86 755-61120055          Email: [email protected]

Website: www.medxing.com

Shenzhen Med-link Electronics Tech Co,.Ltd

AM-806VB REV 

A.1

Revision date: October 24, 2019

Veterinary Temp-pulse Oximeter 

Instructions

 

for

 

u

se

 

AM-806VB

Please read user manual carefully before operate

*

Product picture is for reference only, please refer to the object

Summary of Contents for AM-806VB

Page 1: ...Fax 86 755 61120055 Email kefu medxing com Website www medxing com Shenzhen Med link Electronics Tech Co Ltd AM 806VB REV A 1 Revision date October 24 2019 Veterinary Temp pulse Oximeter Instructions...

Page 2: ...t sites that are not well perfused To prevent this situation please check application sites of animal subject frequently If the initial skin temperature does not exceed 35 C the operation of the senso...

Page 3: ...ing inspection 07 1 12 Packaging 07 Chapter II Operation 08 2 1 Installation and confirmation 08 2 2 Operation and use 10 2 3 Security 11 2 4 Function settings 12 2 5 Menu Settings 13 2 6 List of appl...

Page 4: ...2 Display 7 Bar graph The Pulse Amplitude Indicator 3 Battery indicator 8 Display direction switch 4 Temperature Display 9 Accessory connector 5 Pulse Rate Display bpm 1 4 Technical Specifications Bat...

Page 5: ...rement please use the reusable body cavity temperature probe for animal to measure the same position for at least 90 seconds Medical high frequency signals or defibrillator interference can cause erro...

Page 6: ...the two wavelengths can be calculated thereby the oxygen saturation of the arterial blood can be accurately measured within the range of 70 100 obtaining the pleth waveform and the pulse rate value 1...

Page 7: ...ease contact the Technical Service Department You can contact by Phone 86 755 61120085 Fax 86 755 61120055 Email user07 med linket com 1 11 Unpacking and Inspection 1 12 Included in the package Sectio...

Page 8: ...ning properly or not Caution 2 2 General Operation Preparation 1 Please make sure that the host battery is fully charged If the battery is low please plug in the USB cable to charge 2 Press the power...

Page 9: ...pressure cuff or intravenous injection Make sure the light source is from the internal transparent silica gel Any other light source may cause incorrect SpO2 measurement and venous pulsation Strong l...

Page 10: ...completing the setting the interface will indicate OK Move the cursor to select Exit by short pressing the button and then long press the button to return to the testing interface Setting of SpO2 upp...

Page 11: ...em of Exit to return to the testing interface Return to the testing interface Warning Uncomfortable or painful feeling may appear if using the device ceaselessly especially for animals with blood micr...

Page 12: ...sound which can be turned off by turning off the Alm function Temperature value goes beyond the display range of the device The temperature measurement function is malfunctional If the Alm function is...

Page 13: ...ng time please charge it at least once every 3 months to maintain the battery The best storage environment of the device is between room temperature 10 C and 40 C with humidity not higher than 80 Plea...

Page 14: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 O indicates that the content of the toxic and hazardous substances in all homogeneous materials of the part is subject to the...

Page 15: ......

Page 16: ...stomer name Tel Address Product Name Product REFNO Purchase Date Cityof Purchase FailureDescription Product maintenance card This joint is reserved for customers with this card repair and return Custo...