Technical data
92
Electromagnetic immunity – for equipment and systems that are not life‑supporting
The SONNET 2 is intended for use in the home healthcare environment. The customer
or the user of the SONNET 2 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level Electromagnetic environment – guidance
Conducted RF
IEC 61000‑4‑6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF communications
equipment should be used no closer
than 30 cm to any part of the SONNET 2,
including cables, specified by MED‑EL.
Otherwise degradation of the performance
of the SONNET 2 could result.
6 Vrms in ISM and
amateur radio bands
between 150 kHz and
80 MHz
6 Vrms
Radiated RF
IEC 61000‑4‑3
10 V/m
80 MHz to 2.7 GHz
10 V/m
3 V/m
2.7 GHz to 6 GHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000‑4‑3
27 V/m
380 MHz to 390 MHz
27 V/m
28 V/m
430 MHz to 470 MHz
28 V/m
9 V/m
704 MHz to 787 MHz
9 V/m
28 V/m
800 MHz to 960 MHz
28 V/m
28 V/m
1700 MHz to 1990 MHz
28 V/m
28 V/m
2400 MHz to 2570 MHz
28 V/m
9 V/m
5100 MHz to 5800 MHz
9 V/m
Summary of Contents for Me151 Series
Page 2: ......
Page 3: ...1 CHAPTER 1 Table of contents...
Page 6: ...4...
Page 7: ...5 CHAPTER 2 Introduction...
Page 9: ...7 Intended use Indications Contraindications CHAPTER 3...
Page 13: ...11 SONNET 2 audio processor CHAPTER 4...
Page 48: ...46...
Page 49: ...47 CHAPTER 5 Special considerations for young children...
Page 51: ...49 CHAPTER 6 General precautions and warnings...
Page 60: ...58...
Page 61: ...59 CHAPTER 7 Care and maintenance...
Page 69: ...67 CHAPTER 8 Troubleshooting...
Page 80: ...78...
Page 81: ...79 CHAPTER 9 Technical data...
Page 95: ...93 CHAPTER 10 Appendices...
Page 98: ...96...
Page 99: ......