CHAPTER 2

 

| INSTALLING INTERFACE CARDS 

Drivers and Software 

Refer to the Instructions for installing MED-PC, Delphi, and the hardware drivers, as well as 
running the hardware configuration utility in DOC-010, 

MED-PC IV User’s Manual

. If a PC was 

purchased from MED Associates Inc., the software and drivers will already be installed. If 
necessary, use the Hardware Configuration Utility to set user-defined options for file format, file 
naming, data path, and backup.  

PCI Computer Interface Card   

If the computer was purchased from MED Associates, the PCI card will be factory-installed.  
Otherwise, install the PCI Computer Interface Card into an available full height PCI slot in the 
computer  following the manufacturer’s instructions provided with the computer.  When 
finished, turn the computer on.  The Found New Hardware Wizard should begin automatically; if 
it does not, select 

Start |My Computer | Control Panel | Add New Hardware

.  The drivers are 

located on the MED-PC® installation CD and are automatically installed with MED-PC. 

Decoder Card 

To install the Decoder Card first turn off the power to the interface cabinet (SG-6080D or SG-
6510D).  The decoder card (DIG-700G) should be installed in the leftmost slot of the interface 
cabinet. Connect an SG-210CP two-pin Molex cable from the front of the decoder card to a 28 
VDC power supply in the rear of the cabinet.  (S
ee Figure 2-1.) 

 

 

Figure 2-1 – 28Volt Connection to Decoder Card 

 

  
 

SG-210CP-25 2-Pin Molex Cable 

Summary of Contents for SHUTTLE BOX

Page 1: ...arch SHUTTLE BOX FOR ACTIVE PASSIVE AVOIDANCE MED APA INSTALLATION GUIDE DOC 015 Rev 1 4 Copyright 2013 All Rights Reserved Med Associates Inc P O Box 319 St Albans Vermont 05478 Phone 802 527 2343 Fa...

Page 2: ...MED ASSOCIATES INC MED APA SHUTTTLE BOX PACKAGE i DOC 015 Rev 1 4 Copyright 2013 Med Associates Inc notes...

Page 3: ...1 Introduction 1 Chapter 2 Installing Interface Cards 2 Drivers and Software 2 PCI Computer Interface Card 2 Decoder Card 2 SmartCTRL Interface and Power Supply 3 Chapter 3 Wiring Instructions 5 IR Co...

Page 4: ...y therefore the system requires minimal assembly This manual serves as a guide to complete the wiring of the system Should further assistance be required please visit www med associates com For softwa...

Page 5: ...talled Otherwise install the PCI Computer Interface Card into an available full height PCI slot in the computer following the manufacturer s instructions provided with the computer When finished turn...

Page 6: ...quence starting with Box 1 on the leftmost slot next to the Decoder card Make sure the ports and address offsets are set according to the instructions in DOC 013 SmartCtrl Manual Connect the SmartCTRL...

Page 7: ...4 DOC 015 Rev 1 4 Copyright 2013 Med Associates Inc Connect the DIG 704PCI 2 card to the DIG 700G Decode Card using the DIG 700C Ribbon Cable Figure 2 3 Connect Decoder Card to PCI Card in Computer us...

Page 8: ...e photo beam height to accommodate different sized rats and mice The IR strips connect to an eight channel I R control box ENV 253C to transmit data to MED PC software Connect the outputs from the ENV...

Page 9: ...grid floor using SG 219G 10 cables Please note that there are two ENV 412 Scramblers in the cabinet The Scrambler on the left must connect to the grid floor on the left side of the Shuttle Box and th...

Page 10: ...013 Med Associates Inc Shock and A B Switch Cable Using the SG 219SA cable connect the MED Connection Panel Outputs to the ENV 410B and ENV 415 Switch and Stimulator interface cards Refer to Figure 3...

Page 11: ...connect it to the available output port output 5 is used with standard MedState protocol on the Connection Panel as shown in Figure 4 1 Figure 4 1 Guillotine Door Connection APPENDIX A CONTACT INFORM...