M E D   A S SO C I A T E S   I NC .  

 

E N V - 5 9 7   W AT E R   M A Z E   P O O L   H E AT E R  

- 2 -

 

DOC-213 Rev 1.2 Copyright © 2013 

MED Associates, Inc.

 

CHAPTER 2 | THERMOSTAT SETTINGS   

Setti ng the Temperature  

In order to set the temperature, or make other settings changes, remove the four Phillips screws 
from the corners of the clear protective cover to gain access to the keypad. Be sure your work 
area is clean and dry before removing the cover.  

The upper red display labeled 

PV

 (present value) is the current temperature of the water. The 

lower green display labeled 

SV

 (set value) is your desired temperature.  

To adjust the 

SV

 (set value): 

1.)

 

Press and release the 

SET

 button once, the display will now illuminate all four digits of 

the green 

SV

 display. The “ones” digit will be highlighted and active. 

2.)

 

Press the 

^

 or 

˅

 keys to increase or decrease the temperature in increments of 1˚C. 

3.)

 

Press the 

<r/s

 key and use the 

^

 or 

˅

 keys to increase or decrease the temperature in 

increments of 10˚C. 

4.)

 

Repeat step 3 if you need to adjust the “hundreds” digit value. 

5.)

 

Press and release the 

SET

 key to exit  

6.)

 

Replace the clear protective cover  

 

Changing to ˚F  

In order to change the temperature display to ˚F: 

1.)

 

Press and hold the 

SET

 button until the top red display reads 

Aru

2.)

 

Press  and  release  the 

SET

  button  three  times  until  the  top  red  display  reads 

LCK

.  The 

bottom green display should read 

0011

 indicating the settings menu is locked.  

3.)

 

Use the 

<r/s 

and 

^

 and 

˅

 keys to change the 

LCK

 value to 

1000

 and then press the 

SET 

key. The settings menu is now unlocked and changes can be made. 

4.)

 

Press and hold the 

SET

 key to exit this menu. 

5.)

 

Press and hold the 

SET

 and 

<r/s

 keys simultaneously until the top red display reads 

Cod

.  

6.)

 

Press the 

SET

 key two  times until the top red display reads 

SL2

 

7.)

 

Use the 

<r/s

 and 

^

 and 

˅ 

keys to adjust the value of 

SL2

 to 

0001

 

8.)

 

Press the 

SET

 key 

9.)

 

Press and hold 

SET 

and 

<r/s

 simultaneously to exit the 

Cod 

menu 

10.)

 

Repeat  steps 1-4 to change  the 

LCK

  value  back  to 

0011

  to  prevent  unwanted settings 

changes 

11.)

 

Replace the clear protective cover 