45
TROuBLESHOOTING
8.2
Troubleshooting Quick Guide
Problem
Probable Cause
Solution
When the device is switched
on, the interface language is
not the preferred one.
The desired language has
not been selected.
for the language
selection.
The device does not start
when the switch is positioned
to ON.
The connector on the end
of the power supply cord is
not plugged into the socket
on the back of the device
properly.
Check that the power supply
cord is firmly connected.
The power supply cord is
faulty.
Check that the power supply
socket is functioning.
Replace the power supply
cord.
The fuses are not working.
Replace the fuses
).
The device is switched on but
does not work.
No errors are displayed on
the screen.
The foot pedal plug is not
correctly inserted into the
device socket.
Correctly insert the foot
pedal plug into the socket on
the back of the device.
The foot pedal is not
functioning correctly.
If this occurs during
operation, disconnect the
foot pedal from the device
and use the foot pedal
). After
completing the operation,
contact the Piezosurgery
Inc.’s Service Centre.
The device is switched on but
does not work.
The screen shows an error
message.
for the possible causes
related to the displayed
message.
for the actions to
be taken according to the
message displayed.
When
operating,
the
PIEZOSURGERY
®
medical
handpiece emits a slight
whistle.
The insert is not properly
tightened into the
handpiece.
Unscrew the insert and screw
it back correctly by using the
Mectron torque wrench.
(Refer to
and
)
The irrigation circuit has not
been completely filled.
Fill the irrigation circuit by
using the “pump” button
(Refer to
)