INSTALLATION AND CONFIGuRATION
4.9 Installation of the Inserts
Screw the chosen insert into the
handpiece or extension till the end of the
threading;
WARNING
: Screw onto the
extension only the suitable inserts
supplied in the blister pack. Do not
screw any other type of inserts.
WARNING
: Do not screw the insert
to be adopted with the extension
directly on the handpiece.
WARNING
: PIEZOSURGERY®
inserts can only be installed
onto PIEZOSuRGERY®
handpieces. PIEZOSURGERY® medical
inserts can only be installed onto
PIEZOSURGERY® medical handpieces.
1
Fasten the insert using the PIEZOSURGERY
®
medical or PIEZOSURGERY
®
torque wrench.
To use the Mectron torque wrench
correctly, proceed as follows:
2
Place the insert into the wrench as shown;
CAUTION:
Do not use the
PIEZOSuRGERY® medical L torque
wrench to fasten the inserts. The
PIEZOSuRGERY® medical L torque
wrench must only be used to fasten the
extension.
3