![mectron MICROPIEZO Use And Maintenance Manual Download Page 113](http://html1.mh-extra.com/html/mectron/micropiezo/micropiezo_use-and-maintenance-manual_1759166113.webp)
19
USE
5.2.1.1
Soft Mode
For patients with particularly sensitive teeth,
the distributed power can be automatically
regulated by activating the
SOFT MODE
function .
In
SOFT MODE
, the system will not attempt
to ensure maximum performance during the
treatment, but rather adjust and limit the
power based on the opposing resistance on
the insert, adopting a less invasive reaction.
The
SOFT MODE
can be enabled for all types of
treatment and power levels with the exception
of power level 0.
Furthermore, by using
SOFT MODE
, the operator
has double the number of selectable intensity/
power levels for each type of treatment.
shows the available power levels in
SOFT MODE
.
Treatment
Power Level
Soft Mode
-
0
-
endo
1
Active
1
Non Active
2
Active
2
Non Active
perio - scaling
1
Active
1
Non Active
2
Active
2
Non Active
3
Active
3
Non Active
prosthetics
.
Active
.
Non Active
Table 6 –
Soft Mode
Once selected,
SOFT MODE
remains active even
if the power levels of a specific treatment are
modified, for example, when switching from
perio-scaling
1 to
perio-scaling
2.
The
SOFT MODE
function is deactivated on the
other hand when switching from a power level
of a different treatment, for example, when
switching from
perio-scaling
2 to
endo
1.
5.2.2
Water Delivery
The flow rate of the peristaltic pump can be
adjusted by selecting the numbers along the
left arch
There are 7 flow rate levels:
•
0 : the operation of the pump is off: no
irrigation; flows from the insert;
•
from 1 to 6 : the pump flow rate goes
from 10 ml/min to 33 ml/min ±10%.
NOTE:
The minimum flow rate of the
peristaltic pump can vary from 6 to 12 ml/
min depending on the time of use and state
of wear.
WARNING:
The water flow level 0 can
be selected for all treatments and all power
levels. For the safety of the patient, it is not
recommended to disable the delivery of the
water during the treatments:
perio-scaling
(with any power) e
prosthetics
with inserts
that are not specifically intended for use
without irrigation.
Summary of Contents for MICROPIEZO
Page 2: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti Components Composants...
Page 3: ...MICROPIEZO MANUALE D USO E MANUTENZIONE...
Page 4: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti MICROPIEZO...
Page 8: ...IV MICROPIEZO PAGINA LASCIATA INTENZIONALMENTE BIANCA...
Page 87: ......
Page 88: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Page 89: ...MICROPIEZO USE AND MAINTENANCE MANUAL...
Page 90: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Components MICROPIEZO...
Page 94: ...IV MICROPIEZO PAGE INTENTIONALLY LEFT BLANK...
Page 173: ......
Page 174: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Page 175: ...MICROPIEZO MODE D EMPLOI ET D ENTRETIEN...
Page 176: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Composants MICROPIEZO...
Page 180: ...IV MICROPIEZO PAGE INTENTIONNELLEMENT LAISS E EN BLANC...
Page 259: ......
Page 260: ...02150403 MICROPIEZO V IT EN FR R v 07 du 31 05 2021...
Page 261: ......