![mectron compact piezo P2K Use And Maintenance Manual Download Page 29](http://html1.mh-extra.com/html/mectron/compact-piezo-p2k/compact-piezo-p2k_use-and-maintenance-manual_1759164029.webp)
23
CLEANING AND STERILISATION
Rinse the cavity of the insert with
demineralised water injected under
pressure (3.8 bar) for at least 10 seconds
to eliminate all detergent residue;
10
Place the insert in the ultrasonic tank
immersed in the enzymatic detergent
solution at 40 °C ±2 °C for at least 10
minutes;
11
Remove the insert from the ultrasounds
bath and rinse with demineralized water;
12
Delicately brush the surface of the insert
again using the brush with soft nylon
bristles;